STATE OF THE MARKET - A REVIEW OF AMBULATORY BLOOD-PRESSURE MONITORING DEVICES

被引:99
作者
OBRIEN, E [1 ]
ATKINS, N [1 ]
STAESSEN, J [1 ]
机构
[1] UNIV ZIEKENHUIS GASTHUISBERG, INWENDIGE GENEESKUNDE CARDIOL, KLIN LAB HYPERTENSIE, LOUVAIN, BELGIUM
关键词
BLOOD PRESSURE MONITORING; AMBULATORY; EQUIPMENT DESIGN; EVALUATION STUDIES; DURABLE MEDICAL EQUIPMENT;
D O I
10.1161/01.HYP.26.5.835
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The introduction of 24-hour ambulatory blood pressure measurement into clinical practice created a large market for ambulatory blood pressure measurement devices. Forty-three such devices from 31 manufacturers or suppliers are now available to satisfy a market demand that is likely to increase. The aim of this article is to identify the devices available and then to examine critically any validation studies assessing accuracy and performance. Of the 43 devices available 18 have been validated according to the protocols of the Association for the Advancement of Medical Instrumentation (AAMI) or the British Hypertension Society (BHS) in 25 reported studies. In 9 of these studies the protocol was not adhered to, and the results, which are therefore questionable, are noted but not considered further. Fourteen devices were evaluated according to the accuracy criteria of both protocols, and of these 9 fulfilled the requirements. From this review of 43 devices on the market it may be concluded that: at the time of writing, there is published evidence for only 9 devices meeting the generally accepted AAMI and BHS criteria for accuracy and performance; these are the A&D TM-2420 models 6 and 7 and TM-2421, CH-Druck, Nissei ABPM DS-240, Profilomat, QuietTrak, and SpaceLabs SL-90202 and SL-90207.
引用
收藏
页码:835 / 842
页数:8
相关论文
共 98 条
[11]   BENCH AND AMBULATORY FIELD-EVALUATION OF THE A-AND-D-TM-2420 AUTOMATED SPHYGMOMANOMETER [J].
CONWAY, J .
JOURNAL OF HYPERTENSION, 1991, 9 (06) :577-578
[12]  
COX J, 1992, ANGEING CLIN EXP RES, V4, P85
[13]  
DALLY JFH, 1926, HIGH BLOOD PRESSURE, V13, P33
[14]  
EVANS CE, 1989, J HYPERTENS, V7, P133
[15]  
FITZGERALD DJ, 1982, BRIT HEART J, V48, P572
[16]   INACCURACY OF LONDON SCHOOL OF HYGIENE SPHYGMOMANOMETER [J].
FITZGERALD, DJ ;
OMALLEY, K ;
OBRIEN, ET .
BRITISH MEDICAL JOURNAL, 1982, 284 (6308) :18-19
[17]   EVALUATION OF THE REMLER M2000 BLOOD-PRESSURE RECORDER - COMPARISON WITH INTRAARTERIAL BLOOD-PRESSURE RECORDINGS BOTH AT HOSPITAL AND AT HOME [J].
GOULD, BA ;
HORNUNG, RS ;
KIESO, HA ;
ALTMAN, DG ;
CASHMAN, PMM ;
RAFTERY, EB .
HYPERTENSION, 1984, 6 (02) :209-215
[18]   EVALUATION OF NONINVASIVE BLOOD-PRESSURE MONITORING DEVICES SPACELAB-90202 AND SPACELAB-90207 VERSUS RESTING AND AMBULATORY 24-HOUR INTRAARTERIAL BLOOD-PRESSURE [J].
GROPPELLI, A ;
OMBONI, S ;
PARATI, G ;
MANCIA, G .
HYPERTENSION, 1992, 20 (02) :227-232
[19]   PREVALENCE AND ACCURACY OF HOME SPHYGMOMANOMETERS IN AN URBAN-POPULATION [J].
HAHN, LP ;
FOLSOM, AR ;
SPRAFKA, JM ;
PRINEAS, RJ .
AMERICAN JOURNAL OF PUBLIC HEALTH, 1987, 77 (11) :1459-1461
[20]   24-HOUR AMBULATORY BLOOD-PRESSURE MEASUREMENT IN A PRIMIGRAVID POPULATION [J].
HALLIGAN, A ;
OBRIEN, E ;
OMALLEY, K ;
MEE, F ;
ATKINS, N ;
CONROY, R ;
WALSHE, JJ ;
DARLING, M .
JOURNAL OF HYPERTENSION, 1993, 11 (08) :869-873