ONLINE PH, PARA-CO2 AND PARA-O2 MONITORING-SYSTEM USING 2-WAY CYCLIC PUMPING OF LACTATE RINGER AS A BASE-LINE SOLUTION

被引:6
作者
NAKAMURA, M
YANO, M
SHIBATANI, K
KOMOTO, Y
机构
[1] KURARAY CO LTD, DIV RES & DEV, DEPT BUSINESS & PROD DEV, KITA KU, OSAKA, JAPAN
[2] OKAYAMA UNIV, SCH MED, INST ENVIRONM & DIS, TOTTORI, JAPAN
关键词
D O I
10.1007/BF02442819
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A system for the online, simultaneous and intermittent analysis of blood pH, pCO2 and pO2 using two-way cyclic pumping of lactate Ringer as a baseline solution is described. The multi-gas sensor is composed of a pH-ISFET with a temperature sensor, a pH-ISFET based pCO2 sensor and a Clark-type pO2 sensor, bundled in an epoxy rod, and held in place in a flow cell which is directly connectable to an indwelling cannula. The system is controlled by a microcomputer, which supervises pump operation, temperature-compensation of sensor outputs, one-point calibration using sampled blood analysis, and provides graphical display of monitored results. Results of arterial blood-gas monitoring for mongrel dogs are presented which demonstrate the reliability of the present system over extended periods.
引用
收藏
页码:45 / 50
页数:6
相关论文
共 9 条
[1]  
ASHWOOD ER, 1983, CLIN CHEM, V29, P1877
[2]  
CLARK J, 1985, COMPUT BIOMED RES, V4, P262
[3]   CONTINUOUS MONITORING OF ARTERIAL AND TISSUE PCO2 [J].
KOHAMA, A ;
NAKAMURA, Y ;
NAKAMURA, M ;
YANO, M ;
SHIBATANI, K .
CRITICAL CARE MEDICINE, 1984, 12 (11) :940-942
[4]   CONTINUOUS MONITORING OF INTERSTITIAL FLUID POTASSIUM DURING HEMORRHAGIC-SHOCK IN DOGS [J].
MCKINLEY, BA ;
HOUTCHENS, BA ;
JANATA, J .
CRITICAL CARE MEDICINE, 1981, 9 (12) :845-851
[5]   INVIVO CONTINUOUS MONITORING OF IONIZED CALCIUM IN DOGS USING ION SENSITIVE FIELD-EFFECT TRANSISTORS [J].
MCKINLEY, BA ;
WONG, KC ;
JANATA, J ;
JORDAN, WS ;
WESTENSKOW, DR .
CRITICAL CARE MEDICINE, 1981, 9 (04) :333-339
[6]  
MINDT W, 1973, P JAHRESTAGUNG D GES, P29
[7]   APPLICATION OF CATHETER-TIP ISFET FOR CONTINUOUS INVIVO MEASUREMENT [J].
SHIMADA, K ;
YANO, M ;
SHIBATANI, K ;
KOMOTO, Y ;
ESASHI, M ;
MATSUO, T .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1980, 18 (06) :741-745
[8]  
SHIMADA K, 1980, Patent No. 4218298
[9]   ONLINE PATIENT-MONITORING SYSTEM FOR THE SIMULTANEOUS ANALYSIS OF BLOOD K+, CA2+, NA+ AND PH USING A QUADRUPLE-FUNCTION CHEMFET INTEGRATED-CIRCUIT SENSOR [J].
SIBBALD, A ;
COVINGTON, AK ;
CARTER, RF .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1985, 23 (04) :329-338