A COMPARISON OF SEMIEMPIRICAL MODELS FOR GENERATING TUNGSTEN TARGET X-RAY-SPECTRA

被引:30
作者
BISSONNETTE, JP [1 ]
SCHREINER, LJ [1 ]
机构
[1] MCGILL UNIV, MED PHYS UNIT, MONTREAL H3G 1A4, QUEBEC, CANADA
关键词
D O I
10.1118/1.596848
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Models for the computer generation of tungsten target x-ray spectra proposed by Birch and Marshall [Phys. Med. Biol. 24, 505-517 (1979)] and recently by Tucker, Barnes, and Chakraborty (Med. Phys. 18 211-218 (1991)] are compared. Some basic differences in the equations for the number of bremsstrahlung photons of different energies in the spectra are discussed. The models are compared in terms of their ability to characterize x-ray spectra from constant potential clinical units using three parameter equivalent spectra (EQSPEC) determined from the fit of model generated transmission curves through aluminum to measured data. The Kramers model for x-ray generation is included for completeness. It is shown that two of the models generate very similar x-ray spectra from given transmission curves although the fitting parameters in the EQSPEC characterization differ.
引用
收藏
页码:579 / 582
页数:4
相关论文
共 9 条
[1]   COMPUTATION OF BREMSSTRAHLUNG X-RAY-SPECTRA AND COMPARISON WITH SPECTRA MEASURED WITH A GE(LI) DETECTOR [J].
BIRCH, R ;
MARSHALL, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (03) :505-517
[2]  
BIRCH R, 1979, CATALOGUE SPECTRAL D
[3]   THE 3 PARAMETER EQUIVALENT SPECTRA AS AN INDEX OF BEAM QUALITY [J].
BOONE, JM .
MEDICAL PHYSICS, 1988, 15 (03) :304-310
[4]  
DYSON NA, 1973, XRAYS ATOMIC NUCLEAR
[5]   MEASUREMENTS OF K,L AND M SHELL X-RAY PRODUCTION EFFICIENCIES [J].
GREEN, M ;
COSSLETT, VE .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1968, 1 (04) :425-&
[6]   On the theory of X-ray absorption and of the continuous X- ray spectrum [J].
Kramers, HA .
PHILOSOPHICAL MAGAZINE, 1923, 46 (275) :836-871
[7]   A PARAMETRIZATION OF THE MASS ATTENUATION COEFFICIENTS FOR ELEMENTS WITH Z = 1 TO Z = 92 IN THE PHOTON ENERGY-RANGE FROM APPROXIMATELY 1 TO 150 KEV [J].
OUELLET, RG ;
SCHREINER, LJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) :987-999
[8]   MOLYBDENUM TARGET X-RAY-SPECTRA - A SEMIEMPIRICAL MODEL [J].
TUCKER, DM ;
BARNES, GT ;
WU, XZ .
MEDICAL PHYSICS, 1991, 18 (03) :402-407
[9]   SEMIEMPIRICAL MODEL FOR GENERATING TUNGSTEN TARGET X-RAY-SPECTRA [J].
TUCKER, DM ;