HIGH-RESOLUTION ION ISOLATION WITH THE ION-CYCLOTRON RESONANCE CAPACITIVELY COUPLED OPEN CELL

被引:31
作者
OCONNOR, PB [1 ]
MCLAFFERTY, FW [1 ]
机构
[1] CORNELL UNIV,DEPT CHEM,BAKER LAB CHEM,ITHACA,NY 14853
关键词
D O I
10.1016/1044-0305(95)00232-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
For ion cyclotron resonance, a capacitively coupled open cell variant with fourfold radial symmetry was constructed and tested for axial excitation-ejection of large ions at high resolution. With a reverse of frequency sweep direction, this cell gave substantial improvements in signal-to-noise ratio due to linearization of the excitation electric field. Single isotopic peaks of ubiquitin (8.6-kDa) and carbonic anhydrase (29-kDa) molecular ions could be isolated by selective stored waveform inverse Fourier transform excitation, which yielded an order of magnitude higher isolation resolving power than previously achieved at high mass-to-charge ratio values.
引用
收藏
页码:533 / 535
页数:3
相关论文
共 28 条
[1]   ELIMINATION OF AXIAL EJECTION DURING EXCITATION WITH A CAPACITIVELY COUPLED OPEN TRAPPED-ION CELL FOR FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETRY [J].
BEU, SC ;
LAUDE, DA .
ANALYTICAL CHEMISTRY, 1992, 64 (02) :177-180
[2]   FOURIER-TRANSFORM ELECTROSPRAY INSTRUMENTATION FOR TANDEM HIGH-RESOLUTION MASS-SPECTROMETRY OF LARGE MOLECULES [J].
BEU, SC ;
SENKO, MW ;
QUINN, JP ;
WAMPLER, FM ;
MCLAFFERTY, FW .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1993, 4 (07) :557-565
[3]   IMPROVED FOURIER-TRANSFORM ION-CYCLOTRON-RESONANCE MASS-SPECTROMETRY OF LARGE BIOMOLECULES [J].
BEU, SC ;
SENKO, MW ;
QUINN, JP ;
MCLAFFERTY, FW .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1993, 4 (02) :190-192
[4]   APPLICATION OF MASS-ANALYZED ION KINETIC-ENERGY SPECTROMETRY (MIKES) TO DETERMINATION OF STRUCTURES OF UNKNOWN COMPOUNDS [J].
BOZORGZADEH, MH ;
MORGAN, RP ;
BEYNON, JH .
ANALYST, 1978, 103 (1227) :613-622
[5]   THE INFINITY CELL - A NEW TRAPPED-ION CELL WITH RADIOFREQUENCY COVERED TRAPPING ELECTRODES FOR FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETRY [J].
CARAVATTI, P ;
ALLEMANN, M .
ORGANIC MASS SPECTROMETRY, 1991, 26 (05) :514-518
[6]   STORED WAVE-FORM SIMULTANEOUS MASS-SELECTIVE EJECTION EXCITATION FOR FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETRY [J].
CHEN, L ;
MARSHALL, AG .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1987, 79 (01) :115-125
[7]   PHASE-MODULATED STORED WAVE-FORM INVERSE FOURIER-TRANSFORM EXCITATION FOR TRAPPED ION MASS-SPECTROMETRY [J].
CHEN, L ;
WANG, TCL ;
RICCA, TL ;
MARSHALL, AG .
ANALYTICAL CHEMISTRY, 1987, 59 (03) :449-454
[8]   THE EFFECT OF ION POSITION ON ICR SIGNAL STRENGTH [J].
DUNBAR, RC .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1984, 56 (01) :1-9
[9]   EFFECT OF FREQUENCY SWEEP DIRECTION ON MOTION OF EXCITED IONS IN FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE CELLS [J].
FUJIWARA, M ;
YAMAUCHI, T ;
MIURA, K ;
INOUE, M .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1992, 3 (08) :847-852
[10]   COMPREHENSIVE THEORY OF THE FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE SIGNAL FOR ALL ION TRAP GEOMETRIES [J].
GROSSHANS, PB ;
SHIELDS, PJ ;
MARSHALL, AG .
JOURNAL OF CHEMICAL PHYSICS, 1991, 94 (08) :5341-5352