A DOUBLE PASS SPECTROMETER FOR FAR INFRARED

被引:25
作者
NOLT, IG
KIRBY, RD
LYTLE, CD
SIEVERS, AJ
机构
[1] Laboratory of Atomic and Solid State Physics, Cornell University, Ithaca, NY
来源
APPLIED OPTICS | 1969年 / 8卷 / 02期
关键词
D O I
10.1364/AO.8.000309
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A far ir spectrometer has been built which uses a double pass optical system to increase the spectral dispersion. Because of the energy limitations in this spectral range, this means of increasing the dispersion is an efficient way to improve the resolving power of a grating monochromator. The same optical system is also used to advantage in a lamellar grating interferometer where the increased dispersion improves the performance at higher frequencies. Consideration is also given to instrument errors and signal noise which can affect, and in some cases limit, the present performance of a far ir Fourier interferometer. © 1969 Optical Society of America.
引用
收藏
页码:309 / &
相关论文
共 23 条