QUANTITATIVE CHARACTERIZATION OF THE EDDY-CURRENT FIELDS IN A 40-CM BORE SUPERCONDUCTING MAGNET

被引:35
作者
LIU, Q
HUGHES, DG
ALLEN, PS
机构
[1] UNIV ALBERTA,DEPT PHYS,EDMONTON T6G 2J1,AB,CANADA
[2] UNIV ALBERTA,DEPT APPL SCI MED,EDMONTON,AB,CANADA
关键词
EDDY CURRENT FIELDS; SPATIAL DEPENDENCE;
D O I
10.1002/mrm.1910310112
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The temporal and spatial dependence of the eddy current fields, generated by switching off x, y, and z gradients in a 40-cm bore Bruker superconducting magnet, have been studied by measuring the offset frequency of the proton FID obtained from a small spherical sample. The measurements were made with the pre-emphasis unit deactivated. The data obtained at each location were well fitted to a sum of four exponentially decaying components. The shortest decay is, in each case, associated with the decay of the current in the gradient coil itself, while the other decays are associated with three eddy currents with decay times ranging from 13 to 480 ms. Each of the three eddy currents generates an essentially uniform gradient for at least 4 cm on either side of the isocenter. However, when the eddy currents are generated by either the vertical transverse gradient or the longitudinal gradient, the eddy current gradients are also accompanied by a field shift.
引用
收藏
页码:73 / 76
页数:4
相关论文
共 11 条
[1]   APPLICATION OF INTERLACED ORIENTATION SLICES IN ORTHOGONAL PLANES FOR SYSTEM CHARACTERIZATION AND AS A BASIS FOR STANDARDIZED BRAIN SECTIONS IN MULTIPLE-OBLIQUE POSITIONS [J].
BOESCH, C ;
MARTIN, E .
MAGNETIC RESONANCE IN MEDICINE, 1990, 15 (03) :357-371
[2]   TEMPORAL AND SPATIAL-ANALYSIS OF FIELDS GENERATED BY EDDY CURRENTS IN SUPERCONDUCTING MAGNETS - OPTIMIZATION OF CORRECTIONS AND QUANTITATIVE CHARACTERIZATION OF MAGNET GRADIENT SYSTEMS [J].
BOESCH, C ;
GRUETTER, R ;
MARTIN, E .
MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (02) :268-284
[3]  
DIXON WJ, 1983, BMDP STATISTICAL SOF, P290
[4]   INTENSITY ARTIFACTS IN MRI CAUSED BY GRADIENT SWITCHING IN AN ANIMAL-SIZE NMR MAGNET [J].
HUGHES, DG ;
ROBERTSON, S ;
ALLEN, PS .
MAGNETIC RESONANCE IN MEDICINE, 1992, 25 (01) :167-179
[5]  
HUGHES DG, 1992, 11TH ANN M SOC MAGN, P362
[6]   ANALYTICAL METHOD FOR THE COMPENSATION OF EDDY-CURRENT EFFECTS INDUCED BY PULSED MAGNETIC-FIELD GRADIENTS IN NMR SYSTEMS [J].
JEHENSON, P ;
WESTPHAL, M ;
SCHUFF, N .
JOURNAL OF MAGNETIC RESONANCE, 1990, 90 (02) :264-278
[7]   INVIVO PROTON SPECTROSCOPY IN PRESENCE OF EDDY CURRENTS [J].
KLOSE, U .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (01) :26-30
[8]   THE CORRECTION OF TRANSIENT B0 FIELD SHIFTS FOLLOWING THE APPLICATION OF PULSED GRADIENTS BY PHASE CORRECTION IN THE TIME DOMAIN [J].
ORDIDGE, RJ ;
CRESSHULL, ID .
JOURNAL OF MAGNETIC RESONANCE, 1986, 69 (01) :151-155
[9]   ANALYSIS OF THE TEMPORAL AND SPATIAL DEPENDENCE OF THE EDDY-CURRENT FIELDS IN A 40-CM BORE MAGNET [J].
ROBERTSON, S ;
HUGHES, DG ;
LIU, Q ;
ALLEN, PS .
MAGNETIC RESONANCE IN MEDICINE, 1992, 25 (01) :158-166
[10]   OPTIMIZATION OF EDDY-CURRENT COMPENSATION [J].
VANVAALS, JJ ;
BERGMAN, AH .
JOURNAL OF MAGNETIC RESONANCE, 1990, 90 (01) :52-70