TARGET DIAGNOSTICS FOR INTENSE LITHIUM ION HOHLRAUM EXPERIMENTS ON PARTICLE-BEAM FUSION ACCELERATOR-II

被引:1
作者
LEEPER, RJ
BAILEY, JE
CARLSON, AL
CHANDLER, GA
DERZON, MS
DUKART, RJ
HEBRON, DE
HUNTER, JA
MIX, LP
MOATS, AR
NASH, TJ
OLSON, WR
ROCKETT, PD
RUIZ, CL
TORRES, JA
WENGER, DF
OLSEN, RW
BARBER, TL
LAKE, PW
SCHMIDLAPP, FA
机构
[1] EG&G ENERGY MEASUREMENTS INC,KIRKLAND OPERAT,ALBUQUERQUE,NM 87196
[2] KTECH CORP INC,ALBUQUERQUE,NM 87110
关键词
D O I
10.1063/1.1146334
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A review of the diagnostics used at Sandia National Laboratories to measure the parameters of intense lithium ion-beam hohlraum target experiments on Particle Beam Fusion Accelerator II will be presented. This diagnostic package contains an extensive suite of x-ray spectral and imaging diagnostics that enable measurements of target temperature and x-ray output. The x-ray diagnostics include time-integrated and time-resolved pinhole cameras, energy-resolved one-dimensional streaked imaging diagnostics, time-integrated and time-resolved grazing incidence spectrographs, a transmission grating spectrograph, an elliptical crystal spectrograph, a bolometer array, an 11- element x-ray diode array, and an 11-element PIN diode detector array. The incident Li beam symmetry and an estimate of incident Li beam power density can be measured from ion beam-induced characteristic x-ray line emission and neutron emission. © 1995 American Institute of Physics.
引用
收藏
页码:511 / 514
页数:4
相关论文
共 10 条
[1]  
COOK DL, 1993, 14TH P INT C PLASM P, P63
[2]   PHOTOELECTRIC QUANTUM EFFICIENCIES AND FILTER WINDOW ABSORPTION-COEFFICIENTS FROM 20-EV TO 10-KEV [J].
DAY, RH ;
LEE, P ;
SALOMAN, EB ;
NAGEL, DJ .
JOURNAL OF APPLIED PHYSICS, 1981, 52 (11) :6965-6973
[3]   FAST, LARGE-SIGNAL, FREE-STANDING FOIL BOLOMETER FOR MEASURING ULTRASOFT X-RAY BURST FLUENCE [J].
DEGNAN, JH .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1979, 50 (10) :1223-1226
[4]  
DERZON MS, 1993, P SOC PHOTO-OPT INS, V2002, P31, DOI 10.1117/12.161368
[5]   DEMONSTRATION OF REDUCED SOURCE SIZE BROADENING WITH A JOHANN FOCUSING ELLIPTIC SPECTROGRAPH AND THEORY OF THE 2ND-ORDER SOURCE BROADENING [J].
HAMMEL, BA ;
PHILLION, DW ;
RUGGLES, LE .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (07) :1920-1925
[6]  
HANSON DL, UNPUB USERS GUIDE SN
[7]   INTENSE ION-BEAM DIAGNOSTICS FOR PARTICLE BEAM FUSION EXPERIMENTS ON PBFA-II [J].
LEEPER, RJ ;
STYGAR, WA ;
MAENCHEN, J ;
RUIZ, CL ;
KENSEK, RP ;
JOHNSON, DJ ;
LOCKNER, TR ;
BAILEY, J ;
COOPER, G ;
LEE, JR ;
MEHLHORN, TA ;
MIX, LP ;
STINNETT, RW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (08) :1860-1865
[8]   DEMONSTRATION OF A SOFT-X-RAY AMPLIFIER [J].
MATTHEWS, DL ;
HAGELSTEIN, PL ;
ROSEN, MD ;
ECKART, MJ ;
CEGLIO, NM ;
HAZI, AU ;
MEDECKI, H ;
MACGOWAN, BJ ;
TREBES, JE ;
WHITTEN, BL ;
CAMPBELL, EM ;
HATCHER, CW ;
HAWRYLUK, AM ;
KAUFFMAN, RL ;
PLEASANCE, LD ;
RAMBACH, G ;
SCOFIELD, JH ;
STONE, G ;
WEAVER, TA .
PHYSICAL REVIEW LETTERS, 1985, 54 (02) :110-113
[9]  
MOATS AR, 1995, REV SCI INSTRUM, V66
[10]  
Spielman R.B., 1981, AIP C P, V75, P278