COINCIDENCE SUMMING IN GAMMA-RAY SPECTROSCOPY

被引:93
作者
SEMKOW, TM [1 ]
MEHMOOD, G [1 ]
PAREKH, PP [1 ]
VIRGIL, M [1 ]
机构
[1] SUNY ALBANY,SCH PUBL HLTH,ALBANY,NY 12201
关键词
D O I
10.1016/0168-9002(90)90561-J
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new technique has been developed to calculate coincidence-summing corrections in γ-ray spectroscopy. In this technique the general coincidence-summing equations were derived in matrix notation, which allowed extracting either the first-order correction (combinations of only two coincident γ-rays) or a full correction (all possible combinations of emitted γ-rays). Subsequently, it is shown how the technique can be applied to the determination of the source disintegration rate, γ-ray emission rates or peak efficiencies in the presence of coincidence summing. In particular, the technique has been applied to the determination of peak efficiencies of a germanium detector. The peak efficiencies were iterated self-consistently using the coincidence-summing equations. The above calculation showed that, in general, the full correction is necessary for complicated decay schemes. In addition, a method has been developed to determine the peak-to-total ratio for a germanium detector in the presence of an interfering γ-ray. © 1990.
引用
收藏
页码:437 / 444
页数:8
相关论文
共 16 条
[1]  
Andersson P., 1986, Nuclear Data Sheets, V48, P251, DOI 10.1016/S0090-3752(86)80110-7
[2]  
Andreev D. S., 1972, Instruments and Experimental Techniques, V15, P1358
[3]  
Andreev D.S., 1973, B ACAD SCI USSR PHYS, V37, P41
[4]   DETERMINATION OF THE PHOTON-EMISSION RATES OF THE NBS LONG-LIVED MIXED-RADIONUCLIDE STANDARD [J].
COURSEY, BM ;
HOPPES, DD ;
SCHIMA, FJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 193 (1-2) :1-8
[5]   A NOVEL-APPROACH TO THE DETERMINATION OF IRIDIUM VIA GE-COINCIDENCE NAI(TL)-ANTICOINCIDENCE GAMMA-RAY SPECTROMETRY [J].
CUMMING, JB ;
PAREKH, PP ;
MURALI, AV .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 265 (03) :468-474
[6]   COINCIDENCE SUMMING CORRECTIONS IN GE(LI)-SPECTROMETRY AT LOW SOURCE-TO-DETECTOR DISTANCES [J].
DEBERTIN, K ;
SCHOTZIG, U .
NUCLEAR INSTRUMENTS & METHODS, 1979, 158 (2-3) :471-477
[7]   PRECISE RELATIVE GAMMA-RAY INTENSITIES FOR CALIBRATION OF GE SEMICONDUCTOR-DETECTORS [J].
GEHRKE, RJ ;
HELMER, RG ;
GREENWOOD, RC .
NUCLEAR INSTRUMENTS & METHODS, 1977, 147 (02) :405-423
[8]  
HEATH RL, 1964, US AEC IDO168801 REP, V1
[9]  
Helmer R. G., 1987, Nuclear Data Sheets, V52, P1, DOI 10.1016/S0090-3752(87)80046-7
[10]  
Lederer C.M., 1978, TABLE ISOTOPES, V7th ed.