PRINCIPLES OF STOPPED-FLOW SEQUENTIAL INJECTION-ANALYSIS AND ITS APPLICATION TO THE KINETIC DETERMINATION OF TRACES OF A PROTEOLYTIC-ENZYME

被引:104
作者
RUZICKA, J
GUBELI, T
机构
[1] Department of Chemistry, University of Washington, BG-10, Seattle
关键词
D O I
10.1021/ac00017a006
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The concept of sequential injection analysis is described, and an application of this novel technique to an assay of traces of a proteolytic enzyme is demonstrated. The automated method is based on a stopped-flow procedure with fluorescence detection. Instead of using the conventional constant flow rate, a sinusoidal flow pattern, as generated by a robust cam-driven piston pump, has been applied. The validation of the novel flow injection system using a nonreactive fluorescent compound yielded highly reproducible results over a concentration range of 3 orders of magnitude. The calibration graph for the enzyme activities in the range 5-75-mu-u mL-1 could be fitted a second-order polynomial. The detection limit is 2.5-mu-u mL-1, which corresponds to 7.2 ng mL-1 pure active enzyme.
引用
收藏
页码:1680 / 1685
页数:6