QUANTITATION OF INSULIN INJECTION BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS

被引:31
作者
LOOKABAUGH, M
BISWAS, M
KRULL, IS
机构
[1] MILLIGEN,BURLINGTON,MA 01803
[2] NORTHEASTERN UNIV,DEPT CHEM,BOSTON,MA 02115
[3] NORTHEASTERN UNIV,BARNETT INST,BOSTON,MA 02115
来源
JOURNAL OF CHROMATOGRAPHY | 1991年 / 549卷 / 1-2期
关键词
D O I
10.1016/S0021-9673(00)91446-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
High-performance capillary electrophoresis (HPCE) was evaluated as a potential technique for the regulatory analysis of commerical dosage forms of insulin. A comparison was made to a liquid chromatographic analysis presently being proposed as an official monograph in the United States Pharmacopeia. The salient points of this comparison were accuracy, precision and ease of use. Both authentic (i.e. single blind, spiked) samples and commercial pharmaceutical formulations (injections) were examined. Chromatographic analyses of both commercial formulations and authentic samples were characterized by good precision, with accuracy being supported by results from authentic (spiked) samples. Conventional HPCE (by which is meant a non-micellar electrolyte used with an uncoated, unmodified fused-silica capillary) achieved reasonable accuracy, but less than impressive precision, when applied to authentic samples. When used for commercial formulations, this type of HPCE did not produce a level of accuracy suitable for regulatory purposes, even with the use of an internal standard.
引用
收藏
页码:357 / 366
页数:10
相关论文
共 16 条
[1]  
ANDERSON RL, 1987, PRACTICAL STATISTICS, P76
[2]   DETERMINATION OF INGREDIENTS OF ANTIPYRETIC ANALGESIC PREPARATIONS BY MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY [J].
FUJIWARA, S ;
HONDA, S .
ANALYTICAL CHEMISTRY, 1987, 59 (23) :2773-2776
[3]   DETERMINATION OF CINNAMIC ACID AND ITS ANALOGS BY ELECTROPHORESIS IN A FUSED-SILICA CAPILLARY-TUBE [J].
FUJIWARA, S ;
HONDA, S .
ANALYTICAL CHEMISTRY, 1986, 58 (08) :1811-1814
[4]  
GROSSMAN PD, 1990, AM BIOTECHNOL LAB, V8, P35
[5]   APPLICATION OF FREE-SOLUTION CAPILLARY ELECTROPHORESIS TO THE ANALYTICAL SCALE SEPARATION OF PROTEINS AND PEPTIDES [J].
GROSSMAN, PD ;
COLBURN, JC ;
LAUER, HH ;
NIELSEN, RG ;
RIGGIN, RM ;
SITTAMPALAM, GS ;
RICKARD, EC .
ANALYTICAL CHEMISTRY, 1989, 61 (11) :1186-1194
[6]  
HUANG X, 1989, ANAL CHEM, V61, P776
[7]   BIAS IN QUANTITATIVE CAPILLARY ZONE ELECTROPHORESIS CAUSED BY ELECTROKINETIC SAMPLE INJECTION [J].
HUANG, XH ;
GORDON, MJ ;
ZARE, RN .
ANALYTICAL CHEMISTRY, 1988, 60 (04) :375-377
[8]   ZONE ELECTROPHORESIS IN OPEN-TUBULAR GLASS-CAPILLARIES [J].
JORGENSON, JW ;
LUKACS, KD .
ANALYTICAL CHEMISTRY, 1981, 53 (08) :1298-1302
[9]   CAPILLARY ZONE ELECTROPHORESIS OF PROTEINS IN UNTREATED FUSED-SILICA TUBING [J].
LAUER, HH ;
MCMANIGILL, D .
ANALYTICAL CHEMISTRY, 1986, 58 (01) :166-170
[10]   CONCENTRATION DISTRIBUTIONS IN FREE ZONE ELECTROPHORESIS [J].
MIKKERS, FEP ;
EVERAERTS, FM ;
VERHEGGEN, TPEM .
JOURNAL OF CHROMATOGRAPHY, 1979, 169 (FEB) :1-10