PITFALLS IN SELECTED ION MONITORING IN GAS-CHROMATOGRAPHY MASS-SPECTROMETRY - A THEORETICAL EXAMPLE

被引:9
作者
DEBRABANDER, HF [1 ]
BATJOENS, P [1 ]
VANDENBRAEMBUSSCHE, C [1 ]
DIRINCK, P [1 ]
SMETS, F [1 ]
POTTIE, G [1 ]
机构
[1] INST HYG & EPIDEMIOL,B-1050 BRUSSELS,BELGIUM
关键词
GAS CHROMATOGRAPHY MASS SPECTROMETRY; ANABOLIC STEROIDS; BIOLOGICAL SAMPLES; INTERFERENCES; SELECTED ION MONITORING; STEROIDS;
D O I
10.1016/0003-2670(93)80272-M
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
For the routine determination of residues of growth promoters two important types of low-resolution gas chromatograph-mass spectrometer may be distinguished: the ultra-trace full-scan instrument [e.g., the ion trap mass spectrometer (ITS40)] and most other quadrupole apparatus using the selected ion monitoring (SIM) mode for detecting very small amounts ( < 1-10 ng). In analysing biological extracts interference between matrix components, present at high concentrations, and analytes, present at low concentrations, should be avoided. In this investigation theoretical examples of pitfalls in SIM due to isotope interference (C-13) from matrix components with the analyte were considered. These interferences may lead to false-positive and -negative results and false quantification.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 16 条
  • [11] STEPHANY R W, 1989, Belgian Journal of Food Chemistry and Biotechnology, V44, P139
  • [12] IDENTIFICATION AND CHARACTERIZATION OF 19-NORTESTOSTERONE IN URINE OF MEAT-PRODUCING ANIMALS
    VANDENBROECK, M
    VANVYNCHT, G
    GASPAR, P
    DASNOIS, C
    DELAHAUT, P
    PELZER, G
    DEGRAEVE, J
    MAGHUINROGISTER, G
    [J]. JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1991, 564 (02): : 405 - 412
  • [13] VANPETHEGEM C, COMMUNICATION
  • [14] VERMOESEN A, 1991, J CHROMATOGR-BIOMED, V564, P385
  • [15] 1987, EEC87410 COMM EUR EC
  • [16] 1989, EEC89610 COMM EUR EC