COMPARATIVE-EVALUATION OF DIGITAL RADIOGRAPHY VERSUS CONVENTIONAL RADIOGRAPHY OF FRACTURED SKULLS

被引:15
作者
LANGEN, HJ
KLEIN, HM
WEIN, B
STARGARDT, A
GUNTHER, RW
机构
[1] Department of Diagnostic Radiology, University of Technology Aachen
关键词
DIGITAL LUMINESCENCE RADIOGRAPHY; STORAGE-PHOSPHOR RADIOGRAPHY; RADIATION DOSE; FRACTURE DIAGNOSIS; SKULL;
D O I
10.1097/00004424-199308000-00005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
OBJECTIVES. The authors assessed the relative efficacy of conventional and digital storage-phosphor radiographs for the detection of skull fractures. METHODS. Fifty conventional film-screen radiographs (FSR) and 50 digital storage-phosphor radiographs (DR) with 66 fractures were compared. Five radiologists evaluated image quality and fracture detectability. The results were analyzed by receiver operating characteristic (ROC) curve analysis. RESULTS. With a standard exposure, the ability to evaluate skull fractures was equally good with either technique (ROC area for DR, 0.8954; for FSR, 0.8870). Digital radiography was superior in evaluating nasal bone. For petrosal bone, the DR image simulates an underexposure. This disadvantage compared with FSR can be compensated by image postprocessing. CONCLUSION. In evaluation of skull fractures, radiologists performance with DR is equivalent to FSR.
引用
收藏
页码:686 / 689
页数:4
相关论文
共 15 条
[1]
BICK U, 1991, RADIOLOGE, V31, P8
[2]
BOHNDORF K, 1989, ORTHOPADE, V18, P66
[3]
EXCRETORY UROGRAPHY USING COMPUTED RADIOGRAPHY [J].
FAJARDO, LL ;
HILLMAN, BJ ;
HUNTER, TB ;
CLAYPOOL, HR ;
WESTERMAN, BR ;
MOCKBEE, B .
RADIOLOGY, 1987, 162 (02) :345-351
[4]
GALANSKI M, 1986, RADIOLOGISCHE DIAGNO, V5, P104
[5]
FRACTURE DIAGNOSIS USING THE DIGITAL STORAGE PHOSPHORUS SYSTEM [J].
KLEIN, HM ;
WEIN, B ;
LANGEN, HJ ;
GLASER, KH ;
STARGARDT, A ;
GUNTHER, RW .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1991, 154 (06) :582-586
[6]
KRUG B, 1990, FORTSCHR RONTG NEUEN, V152, P131
[7]
CONVENTIONAL X-RAY-EXAMINATION AND COMPUTED-TOMOGRAPHY IN ORBITAL FLOOR FRACTURES [J].
LANGEN, HJ ;
DAUS, HJ ;
BOHNDORF, K ;
KLOSE, K .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NUKLEARMEDIZIN, 1989, 150 (05) :582-587
[8]
METZ CE, 1986, INVEST RADIOL, V21, P720, DOI 10.1097/00004424-198609000-00009
[9]
DIGITAL LUMINESCENCE RADIOGRAPHY (DLR) IN THE DIAGNOSIS OF TRAUMATIC BONE-LESIONS [J].
MULLER, RD ;
BUDDENBROCK, B ;
KOCK, HJ ;
VOSS, M ;
JOHN, V ;
BOSCH, E ;
SCHMIDT, G ;
HIRCHE, H ;
LOHR, E ;
SCHMITNEUERBURG, KP .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1991, 154 (06) :575-581
[10]
NONDISPLACED FRACTURES - SPATIAL-RESOLUTION REQUIREMENTS FOR DETECTION WITH DIGITAL SKELETAL IMAGING [J].
MURPHEY, MD ;
BRAMBLE, JM ;
COOK, LT ;
MARTIN, NL ;
DWYER, SJ .
RADIOLOGY, 1990, 174 (03) :865-870