POLARIZATION PROPERTIES OF SMALL-SURFACE-AREA PACEMAKER ELECTRODES - IMPLICATIONS ON RELIABILITY OF SENSING AND PACING

被引:19
作者
FISCHLER, H [1 ]
机构
[1] UNIV PENN,DEPT BIOENGN,PHILADELPHIA,PA 19174
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1979年 / 2卷 / 04期
关键词
cardiac pacing; electrode polarization; pacemaker electrodes;
D O I
10.1111/j.1540-8159.1979.tb05216.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
FISCHLER, H. Polarization properties of small‐surfece‐area pacemaker electrodes—implications on reliability of sensing and pacing. The polarization characteristics of Pt‐Jr and Elgiloy small‐surface‐area (10–12.5 mm2) pacemaker electrodes were studied at AC linear (sensing) and DC non‐linear (pacing) conditions. The electrodes’ AC polarization impedance was approximately equal to the demand pacemaker's input impedance, which causes waveform distortions of the sensed R‐wave potentials. The pacemaker's coupling capacitor adds to the distortion's effect. As a result of amplitude attenuation (up to ∼50%) and slew rate changes, the pacemaker may fail to recognize the ventricular complexes, reverting to hazardous competitive pacing. The impact of the DC polarization elements and of the coupling capacitor on the effectiveness of pacing was examined. The deficiency of small area electrodes was pointed out, this being counter‐balanced to some extent by their lower pacing thresholds. The necessity to ensure a sufficient safety margin between the pacing pulse and the stimulation threshold, to avoid possible reasons for creation of exit blocks, was stressed. In reducing the electrode surface, it is advisable not to step down below a 15–20 mm2 effective area for Pt‐Fr, and a 40–50 mm2 area for Elgiloy. Copyright © 1979, Wiley Blackwell. All rights reserved
引用
收藏
页码:403 / 416
页数:14
相关论文
共 25 条
[11]   POLARIZATION PHENOMENA RELATING TO PHYSIOLOGICAL ELECTRODES [J].
GREATBATCH, W ;
PIERSMA, B ;
SHANNON, FD ;
CALHOON, SW .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1969, 167 (A2) :722-+
[12]   MYCOARDIAL AND ENDOCARDIAC ELECTRODES FOR CHRONIC IMPLANTATION [J].
GREATBATCH, W ;
CHARDACK, WM .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1968, 148 (A1) :234-+
[13]   FAILURE OF DEMAND PACING WITH SMALL SURFACE-AREA ELECTRODES [J].
HUGHES, HC ;
BROWNLEE, RR ;
TYERS, GFO .
CIRCULATION, 1976, 54 (01) :128-132
[14]   NONLINEARITY OF CARDIAC PACEMAKER ELECTRODES [J].
JARON, D ;
BRILLER, SA ;
SCHWAN, HP ;
GESELOWITZ, DB .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1969, BM16 (02) :132-+
[15]   A MATHEMATICAL MODEL FOR POLARIZATION IMPEDANCE OF CARDIAC PACEMAKER ELECTRODES [J].
JARON, D ;
SCHWAN, HP ;
GESELOWI.DB .
MEDICAL & BIOLOGICAL ENGINEERING, 1968, 6 (06) :579-&
[16]  
MILLMAN J, 1956, PULSE DIGITAL CIRCUI, P36
[17]   CHARACTERISTICS OF INTRA-CARDIAC ELECTROGRAMS [J].
MYERS, GH ;
KRESH, YM ;
PARSONNET, V .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1978, 1 (01) :90-103
[18]  
OHM O, 1978, 1ST EUR S CARD PAC L, V79
[19]  
PARSONNET V, 1977, 5TH INT S CARD PAC T, P501
[20]   EQUIVALENT DIAGRAM FOR INTERFACE IMPEDANCE OF METAL NEEDLE ELECTRODES [J].
POLLAK, V .