TECHNIQUE FOR SIMULTANEOUS DUAL ENERGY SCANNING
被引:24
作者:
RITCHINGS, RT
PULLAN, BR
机构:
[1] Department of Medical Biophysics, Stopford Medical School, University of Manchester, Manchester, M13 9PT, Oxford Road
关键词:
Attenuation values;
Computed tomography;
Dual energy;
D O I:
10.1097/00004728-197912000-00028
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
A method of deriving simultaneous dual energy scans on a conventional EMI CT5005 scanner is described. The only change that has to be made to the scanner involves the detector collimator where alternate slits are covered by a thin metal foil. The detector readings from the covered and uncov ered slits are separated by software before picture processing to give two pictures from one scan—one at the usual X-ray energy and the other at a higher X-ray energy due to the beam hardening produced by the foil. A procedure is then outlined for converting these into pictures showing the effective electron density and the effective atomic number for the whole scan field. © 1979 Raven Press, New York.