FREQUENCY-RESPONSE OF PVDF NEEDLE-TYPE HYDROPHONES

被引:43
作者
FAY, B
LUDWIG, G
LANKJAER, C
LEWIN, PA
机构
[1] DREXEL UNIV,INST SCI,PHILADELPHIA,PA 19104
[2] DREXEL UNIV,DEPT ELECT & COMP ENGN & BIOMED ENGN,PHILADELPHIA,PA 19104
[3] FORCES INST,DK-2605 BRONDBY,DENMARK
关键词
ULTRASONIC PVDF POLYMER HYDROPHONE PROBE; PVDF NEEDLE HYDROPHONE; ULTRASONIC HYDROPHONE CALIBRATION;
D O I
10.1016/0301-5629(94)90004-3
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This paper examines the factors governing the frequency response of ultrasonic polyvinylidene flouride (PVDF) polymer needle-type hydrophones, in particular the sensitivity variations in the lower frequency range of 1-6 MHz. A theoretical model was used to analyze the influence of the hydrophone's diameter, the metal electrodes, thickness, PVDF material properties, the adhesive layer acoustical characteristics and the backing material, on the frequency response of the hydrophone. The results of the theoretical modelling differ by less than +/- 0.5 dB from those obtained experimentally from the reciprocity calibration in the frequency range 1-20 MHz. It is shown that the needle hydrophone's diameter and backing material are the main reasons for the sensitivity variations observed in the frequency range below 6 MHz.
引用
收藏
页码:361 / 366
页数:6
相关论文
共 22 条
[11]  
LEWIN PA, 1993, ULTRASONIC EXPOSIMET, P186
[12]  
LEWIN PA, 1979, 79 ULTR INT P NEW YO, P279
[13]  
LEWIS PA, 1988, APPLICATIONS FERROEL, P162
[14]   CALIBRATION OF HYDROPHONES BASED ON RECIPROCITY AND TIME-DELAY SPECTROMETRY [J].
LUDWIG, G ;
BRENDEL, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (02) :168-174
[15]   INTERLABORATORY COMPARISON OF HYDROPHONE CALIBRATIONS [J].
PRESTON, RC ;
BACON, DR ;
CORBETT, SS ;
HARRIS, GR ;
LEWIN, PA ;
MCGREGOR, JA ;
OBRIEN, WD ;
SZABO, TL .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (02) :206-213
[16]  
ROBINSON SP, 1988, PHYSICS MED ULTRASOU, V2, P79
[17]   A COMPUTERIZED SYSTEM FOR MEASURING THE ACOUSTIC OUTPUT FROM DIAGNOSTIC ULTRASOUND EQUIPMENT [J].
SCHAFER, ME ;
LEWIN, PA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (02) :102-109
[18]  
SCHAFER ME, 1993, ULTRASONIC EXPOSIMET, P218
[19]  
SHOMBERT DG, 1984, MED PHYS, V9, P484
[20]  
1986, IEC866 INT EL COMM D