DETERMINATION OF CADMIUM IN URINE BY ELECTRO-THERMAL ATOMIC-ABSORPTION SPECTROMETRY

被引:29
作者
CARMACK, GD [1 ]
EVENSON, MA [1 ]
机构
[1] UNIV WISCONSIN,DEPT MED,MADISON,WI 53706
关键词
D O I
10.1021/ac50043a029
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple procedure for the direct, electrothermal AAS determination of cadmium in urine and other materials containing high-salt concentrations is described. Optical, temperature-controlled heating of the furnace allows cadmium to be selectively atomized from the matrix at 900 °C. Matrix modification, including impregnating the carbon tube with molybdenum, further reduces the background absorption. The method is rapid, has minimal matrix interferences, and requires no sample preparation. The limit of detection is 0.05 μg/L with an average precision of 4%. © 1979, American Chemical Society. All rights reserved.
引用
收藏
页码:907 / 911
页数:5
相关论文
共 37 条
[11]   APPLICATION OF MUNICIPAL REFUSE AND LIQUID SEWAGE SLUDGE TO AGRICULTURAL LAND .1. FIELD STUDY [J].
KING, LD ;
RUDGERS, LA ;
WEBBER, LR .
JOURNAL OF ENVIRONMENTAL QUALITY, 1974, 3 (04) :361-366
[12]   COMPARATIVE STUDY ON METHODS FOR CADMIUM ANALYSIS OF GRAIN WITH AN APPLICATION TO POLLUTION EVALUATION [J].
KJELLSTROM, T ;
LIND, B ;
LINNMAN, L ;
NORDBERG, G .
ENVIRONMENTAL RESEARCH, 1974, 8 (01) :92-106
[13]   TEMPERATURE CONTROLLED HEATING OF GRAPHITE TUBE ATOMIZER IN FLAMELESS ATOMIC-ABSORPTION SPECTROMETRY [J].
LUNDGREN, G ;
LUNDMARK, L ;
JOHANSSO.G .
ANALYTICAL CHEMISTRY, 1974, 46 (08) :1028-1031
[14]  
Manning D.C., 1976, ATOM ABSORPT NEWSL, V15, P42
[15]  
MANNING DC, 1978, ATOM ABSORPT NEWSL, V17, P43
[16]  
MATSON WR, 1971, TRACE SUBSTANCES ENV, V4, P396
[17]  
NATELSON S, 1975, PRINCIPLES APPLIED C
[18]  
NORDBERG G F, 1972, Environmental Physiology and Biochemistry, V2, P7
[19]  
Nordberg G. F, 1976, EFFECTS DOSE RESPONS, pI
[20]  
NORMIYAMA K, 1976, EFFECTS DOSE RESPONS, P372