A NEW PLASTIC EVACUATED TUBE WITH PLASMA SEPARATOR

被引:14
作者
LANDT, M
WILHITE, TR
SMITH, CH
机构
[1] WASHINGTON UNIV,SCH MED,DEPT PATHOL,ST LOUIS,MO 63110
[2] ST LOUIS CHILDRENS HOSP,ST LOUIS,MO
关键词
TUBE; PLASMA SEPARATOR; CENTRIFUGATION;
D O I
10.1002/jcla.1860090205
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A new plastic-walled, evacuated blood collection tube (Terumo Venoject II) with plasma separator was evaluated for characteristics related to the use of plastic in place of glass. Plastic tubes were highly resistant to breakage through handling mishaps or failure during centrifugation. Higher centrifugation speed (13,600g) was well tolerated, and centrifugation times as short as 3 minutes at 13,6009 effectively cleared plasma of cellular components, at high plasma yield. Plastic tubes were nearly completely combustible under incineration conditions commonty used for medical waste, and plastic tubes effectively retained Vacuum during the typical shelf life of evacuated blood collection tubes. Collection of carbamezepine specimens in plastic tubes decreased levels an average of 6.8% compared to a heparinized glass tube control under conditions approximating routine use; levels of other drugs (phenytoin, phenobarbital, valproic acid, theophylline, and cyclosporine) were less significantly affected. This modest decrease appeared to result from a combination of an immediate interaction with the plastic surface of the tubes and a time-dependent interaction with the olefin-based separator. Modest but clear benefits, including reduction in specimen breakage, reduced centrifugation time and reduced solid waste after incineration derive from the use of plastic in place of glass in evacuated brood collection tubes. (C) 1995 Wiley-Liss, Inc.
引用
收藏
页码:101 / 106
页数:6
相关论文
共 9 条
[1]  
BERGQVIST Y, 1984, CLIN CHEM, V30, P465
[2]  
CALAM RR, 1988, J CLIN IMMUNOASSAY, V11, P86
[3]   EVALUATION OF THE GREINER PLASMA SEPARATOR BLOOD TUBE [J].
FAULKNER, AM ;
LUKESHALL, AM ;
WHITE, GW .
ANNALS OF CLINICAL BIOCHEMISTRY, 1990, 27 :386-387
[4]  
HILL BM, 1992, CLIN CHEM, V38, P1474
[5]   SUITABILITY OF COLLECTION TUBES WITH SEPARATOR GELS FOR THERAPEUTIC DRUG-MONITORING [J].
KOCH, TR ;
PLATOFF, G .
THERAPEUTIC DRUG MONITORING, 1990, 12 (03) :277-280
[6]  
LAESSIG RH, 1976, AM J CLIN PATHOL, V66, P639
[7]  
LANDT M, 1993, CLIN CHEM, V39, P1712
[8]   STABILITY OF PHENYTOIN IN BLOOD COLLECTED IN VACUUM BLOOD COLLECTION TUBES [J].
PARISH, RC ;
ALEXANDER, T .
THERAPEUTIC DRUG MONITORING, 1990, 12 (01) :85-89
[9]   EFFECT OF SERUM SEPARATOR BLOOD COLLECTION TUBES ON DRUG CONCENTRATIONS [J].
QUATTROCCHI, F ;
KARNES, HT ;
ROBINSON, JD ;
HENDELES, L .