OPTIMAL-CONTROL SYSTEM FOR THE INTRAAORTIC BALLOON PUMP

被引:11
作者
SMITH, B
BARNEA, O
MOORE, TW
JARON, D
机构
[1] DREXEL UNIV,INST BIOMED ENGN & SCI,PHILADELPHIA,PA 19104
[2] TEL AVIV UNIV,SCH ENGN,BIOMED ENGN PROGRAM,IL-69978 TEL AVIV,ISRAEL
关键词
ALGORITHM; CONTROL; INTRAAORTIC BALLOON PUMP; OPTIMIZATION;
D O I
10.1007/BF02447105
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
An optimal control system for the intra-aortic balloon pump (IABP) is presented. Control of the IABP is based on a performance index formulated to reflect a tradeoff between maximising cardiac oxygen supply and minimising cardiac oxygen consumption. In the performance index, mean diastolic pressure (MDP) was used to represent oxygen availability and peak systolic pressure (PSP) was used to represent oxygen consumption. An algorithm, implemented using an 8-bit microcomputer, changes the deflation time of the IABP to maximise this performance index by using an optimisation technique that employs both a search and an approximation. The search produces three equally spaced points which define a region that includes the maximum of the performance index. From these points, the optimum deflation time is estimated by a quadratic approximation. The algorithm has been successfully tested using performance index curves generated by computer simulations.
引用
收藏
页码:180 / 184
页数:5
相关论文
共 11 条
[1]  
ADBY PR, 1974, INTRO OPTIMIZATION M
[2]  
BARNEA O, 1987, THESIS DREXEL U PHIL
[3]  
BARNEA O, 1989, 11TH INT C IEEE EMBS
[4]  
BOX MJ, 1969, NONLINEAR OPTIMIZATI
[5]   FEASIBILITY OF CLOSED-LOOP CONTROL OF INTRA-AORTIC BALLOON PUMPING [J].
CLARK, JW ;
KANE, GR ;
BOURLAND, HM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (06) :404-412
[6]  
JABLKOWSKI, 1980, ISAO IFAC S, V2, P245
[7]   THEORETICAL CONSIDERATIONS REGARDING THE OPTIMIZATION OF CARDIAC ASSISTANCE BY INTRAAORTIC BALLOON PUMPING [J].
JARON, D ;
MOORE, TW ;
HE, P .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (03) :177-186
[8]   CONTROL OF INTRAAORTIC BALLOON PUMPING - THEORY AND GUIDELINES FOR CLINICAL-APPLICATIONS [J].
JARON, D ;
MOORE, TW ;
HE, P .
ANNALS OF BIOMEDICAL ENGINEERING, 1985, 13 (02) :155-175
[9]  
KANE GR, 1971, T AM SOC ART INT ORG, V17, P148
[10]   OPTIMAL QRS DETECTOR [J].
THAKOR, NV ;
WEBSTER, JG ;
TOMPKINS, WJ .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1983, 21 (03) :343-350