TEMPERATURE DISTRIBUTIONS DURING CLINICAL SCANNED, FOCUSED ULTRASOUND HYPERTHERMIA TREATMENTS

被引:41
作者
HYNYNEN, K
SHIMM, D
ANHALT, D
STEA, B
SYKES, H
CASSADY, JR
ROEMER, RB
机构
[1] UNIV ARIZONA, DEPT AEROSP & MECH ENGN, TUCSON, AZ 85724 USA
[2] ARIZONA CANC CTR, TUCSON, AZ 85724 USA
关键词
D O I
10.3109/02656739009140971
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
In this study a scanned focused ultrasound (SFUS) system was used to heat 66 tumours at various anatomical locations in 52 patients. A total of 160 treatments were given. On average, temperatures were measured in 14 or 15 locations in the scanned volume. The time-averaged temperatures over the 30 min treatment period in the best treatment of each tumour were 44.0 ± 2.4°C (mean ± SD) and 39.6 ± 1.5°C at the location of the highest and lowest sensor, respectively. On average, 39% of the sensors were above 42.5°C. When only the cases that were judged to be good candidates for the hyperthermia device were analysed, 64% of the sensors reached a temperature over 42.5°C with the highest temperature achieved being 45.9±2.3°C and the lowest 40.7 ± 1.4°C. Although the system tested has many technical limitations (for example, fixed frequency, beam geometry and power during the scan cycle), the results demonstrate that therapeutic temperatures can be achieved in many tumours. Significantly better temperatures are expected when all of the theoretical potential of scanned focused ultrasound systems has been used. © 1990 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.
引用
收藏
页码:891 / 908
页数:18
相关论文
共 25 条
  • [1] ULTRASOUND-INDUCED HYPERTHERMIA FOR THE TREATMENT OF HUMAN SUPERFICIAL TUMORS
    CORRY, PM
    BARLOGIE, B
    TILCHEN, EJ
    ARMOUR, EP
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1982, 8 (07): : 1225 - 1229
  • [2] EVALUATION OF EQUIPMENT FOR HYPERTHERMIC TREATMENT OF CANCER
    CORRY, PM
    JABBOURY, K
    KONG, JS
    ARMOUR, EP
    MCCRAW, FJ
    LEDUC, T
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (01) : 53 - 74
  • [3] DEWHIRST MW, 1982, MED PHYS MONOGR, V8, P495
  • [4] THERMAL CONDUCTION ERRORS OF MANGANIN-CONSTANTAN THERMOCOUPLE ARRAYS
    DICKINSON, RJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (05) : 445 - 453
  • [5] EXPERIENCE WITH A MULTITRANSDUCER ULTRASOUND SYSTEM FOR LOCALIZED HYPERTHERMIA OF DEEP TISSUES
    FESSENDEN, P
    LEE, ER
    ANDERSON, TL
    STROHBEHN, JW
    MEYER, JL
    SAMULSKI, TV
    MARMOR, JB
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (01) : 126 - 135
  • [6] A SCANNED, FOCUSED, MULTIPLE TRANSDUCER ULTRASONIC SYSTEM FOR LOCALIZED HYPERTHERMIA TREATMENTS
    HYNYNEN, K
    ROEMER, R
    ANHALT, D
    JOHNSON, C
    XU, ZX
    SWINDELL, W
    CETAS, T
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1987, 3 (01) : 21 - 35
  • [7] TEMPERATURE ELEVATION AT MUSCLE BONE INTERFACE DURING SCANNED, FOCUSED ULTRASOUND HYPERTHERMIA
    HYNYNEN, K
    DEYOUNG, D
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (03) : 267 - 279
  • [8] TEMPERATURE-MEASUREMENTS DURING ULTRASOUND HYPERTHERMIA
    HYNYNEN, K
    EDWARDS, DK
    [J]. MEDICAL PHYSICS, 1989, 16 (04) : 618 - 626
  • [9] THE EFFECT OF BLOOD PERFUSION RATE ON THE TEMPERATURE DISTRIBUTIONS INDUCED BY MULTIPLE, SCANNED AND FOCUSED ULTRASONIC BEAMS IN DOGS KIDNEYS INVIVO
    HYNYNEN, K
    DEYOUNG, D
    KUNDRAT, M
    MOROS, E
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1989, 5 (04) : 485 - 497
  • [10] THE EFFECT OF SCANNING SPEED ON TEMPERATURE AND EQUIVALENT THERMAL EXPOSURE DISTRIBUTIONS DURING ULTRASOUND HYPERTHERMIA INVIVO
    HYNYNEN, K
    ROEMER, R
    MOROS, E
    JOHNSON, C
    ANHALT, D
    [J]. IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1986, 34 (05) : 552 - 559