VALIDATION OF A DOSE-POINT KERNEL CONVOLUTION TECHNIQUE FOR INTERNAL DOSIMETRY

被引:62
作者
GIAP, HB
MACEY, DJ
BAYOUTH, JE
BOYER, AL
机构
[1] Dept. of Radiat. Phys., Texas Univ., Houston, TX
关键词
D O I
10.1088/0031-9155/40/3/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this study was to validate a dose-point kernel convolution technique that provides a three-dimensional (3D) distribution of absorbed dose from a 3D distribution of the radionuclide I-131. A dose-point kernel for the penetrating radiations was calculated by a Monte Carlo simulation and cast in a 3D rectangular matrix. This matrix was convolved with the 3D activity map furnished by quantitative single-photon-emission computed tomography (SPECT) to provide a 3D distribution of absorbed dose. The convolution calculation was performed using a 3D fast Fourier transform (FFT) technique, which takes less than 40 s for a 128 x 128 x 16 matrix on an Intel 486 DX2 (66 MHz) personal computer. The calculated photon absorbed dose was compared with values measured by thermoluminescent dosimeters (TLDs) inserted along the diameter of a 22 cm diameter annular source of I-131. The mean and standard deviation of the percentage difference between the measurements and the calculations were equal to -1% and 3.6%, respectively. This convolution method was also used to calculate the 3D dose distribution in an Alderson abdominal phantom containing a liver, a spleen, and a spherical tumour volume loaded with various concentrations of I-131. By averaging the dose calculated throughout the liver, spleen, and tumour the dose-point kernel approach was compared with values derived using the MIRD formalism, and found to agree to better than 15%.
引用
收藏
页码:365 / 381
页数:17
相关论文
共 33 条
[1]  
BERGER MJ, 1978, MIRD2 SOC NUCL MED P
[2]  
BERGER MJ, 1978, MIRD7 SOC NUCL MED P
[3]  
Berger MJ., 1973, ORNL DOCUMENTATION R
[4]   BRACHYTHERAPY SEED DOSE DISTRIBUTION CALCULATION EMPLOYING THE FAST FOURIER-TRANSFORM [J].
BOYER, AL ;
MOK, EC .
MEDICAL PHYSICS, 1986, 13 (04) :525-529
[5]  
BUCHEGGER F, 1991, J NUCL MED, V32, P1414
[8]   REQUIREMENTS FOR A TREATMENT PLANNING SYSTEM FOR RADIOIMMUNOTHERAPY [J].
DENARDO, GL ;
RAVENTOS, A ;
HINES, HH ;
SCHEIBE, PO ;
MACEY, DJ ;
HAYS, MT ;
DENARDO, SJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (02) :335-348
[9]  
ELLETT WH, 1978, MIRD8 SOC NUCL MED P
[10]   RADIOIMMUNOTHERAPY OF ATHYMIC MICE BEARING HUMAN-COLON CARCINOMAS WITH MONOCLONAL-ANTIBODY B72.3 - HISTOLOGICAL AND AUTORADIOGRAPHIC STUDY OF EFFECTS ON TUMORS AND NORMAL ORGANS [J].
ESTEBAN, JM ;
SCHLOM, J ;
MORNEX, F ;
COLCHER, D .
EUROPEAN JOURNAL OF CANCER & CLINICAL ONCOLOGY, 1987, 23 (06) :643-655