VALIDATION OF A DOSE-POINT KERNEL CONVOLUTION TECHNIQUE FOR INTERNAL DOSIMETRY

被引:62
作者
GIAP, HB
MACEY, DJ
BAYOUTH, JE
BOYER, AL
机构
[1] Dept. of Radiat. Phys., Texas Univ., Houston, TX
关键词
D O I
10.1088/0031-9155/40/3/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this study was to validate a dose-point kernel convolution technique that provides a three-dimensional (3D) distribution of absorbed dose from a 3D distribution of the radionuclide I-131. A dose-point kernel for the penetrating radiations was calculated by a Monte Carlo simulation and cast in a 3D rectangular matrix. This matrix was convolved with the 3D activity map furnished by quantitative single-photon-emission computed tomography (SPECT) to provide a 3D distribution of absorbed dose. The convolution calculation was performed using a 3D fast Fourier transform (FFT) technique, which takes less than 40 s for a 128 x 128 x 16 matrix on an Intel 486 DX2 (66 MHz) personal computer. The calculated photon absorbed dose was compared with values measured by thermoluminescent dosimeters (TLDs) inserted along the diameter of a 22 cm diameter annular source of I-131. The mean and standard deviation of the percentage difference between the measurements and the calculations were equal to -1% and 3.6%, respectively. This convolution method was also used to calculate the 3D dose distribution in an Alderson abdominal phantom containing a liver, a spleen, and a spherical tumour volume loaded with various concentrations of I-131. By averaging the dose calculated throughout the liver, spleen, and tumour the dose-point kernel approach was compared with values derived using the MIRD formalism, and found to agree to better than 15%.
引用
收藏
页码:365 / 381
页数:17
相关论文
共 33 条
[11]  
GIAP HB, 1994, J NUCL MED, V35, pP123
[12]  
GIAP HB, 1993, MED PHYS, V20, P909
[13]  
GIAP HB, 1993, MED PHYS, V20, P865
[14]  
GRIFFITH MH, 1988, J NUCL MED, V29, P1795
[16]  
HUMM JL, 1990, J NUCL MED, V31, P75
[17]   PHOTON CONTRIBUTION TO TUMOR DOSE FROM CONSIDERATIONS OF I-131 RADIOLABELED ANTIBODY UPTAKE IN LIVER, SPLEEN, AND WHOLE-BODY [J].
JOHNSON, TK ;
COLBY, SB .
MEDICAL PHYSICS, 1993, 20 (06) :1667-1674
[18]  
JONES PL, 1986, CANCER IMMUNOL IMMUN, V22, P139
[19]   CALCULATION OF RADIATION-DOSES FOR NONUNIFORMLY DISTRIBUTED BETA-RADIONUCLIDES AND GAMMA-RADIONUCLIDES IN SOFT-TISSUE [J].
KWOK, CS ;
PRESTWICH, WV ;
WILSON, BC .
MEDICAL PHYSICS, 1985, 12 (04) :405-412
[20]   EFFECT OF TISSUE INHOMOGENEITY ON DOSE DISTRIBUTION OF POINT SOURCES OF LOW-ENERGY ELECTRONS [J].
KWOK, CS ;
BIALOBZYSKI, PJ ;
YU, SK ;
PRESTWICH, WV .
MEDICAL PHYSICS, 1990, 17 (05) :786-793