A VERSATILE SAMPLE INTRODUCTION SYSTEM WITH GRAPHITE TUBE FURNACE FOR INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY

被引:18
作者
KUMAMARU, T
OKAMOTO, Y
MATSUO, H
机构
[1] Hiroshima Univ, Hiroshima, Jpn, Hiroshima Univ, Hiroshima, Jpn
关键词
SPECTROSCOPY; EMISSION; -; Accessories;
D O I
10.1366/0003702874448210
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This note details an adaptation of a commercially available GFA to ICPAES that retained as much of the original design of the atomizer as possible, so that the apparatus could be switched easily between conventional GFA-AAS and GFV-ICPAES. The described modifications of the GFA were undertaken with two aims: to minimize the cost of conversion from GFA-AAS to GFV-ICPAES and to optimize the efficiency of introducing analyte vapor into the ICP. The recommended procedure for the operation of the GFV device is presented. A comparison of detection limits between the continuous pneumatic nebulization and the GFV sample introduction has been made. The modifications necessary to utilize the GFV system were straightforward, and the system could be switched between ICPAES and AAS systems in a few minutes.
引用
收藏
页码:918 / 920
页数:3
相关论文
共 6 条
[1]   ANALYSIS OF MICROAMOUNTS OF BIOLOGICAL SAMPLES BY EVAPORATION IN A GRAPHITE-FURNACE AND INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROSCOPY [J].
AZIZ, A ;
BROEKAERT, JAC ;
LEIS, F .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (05) :369-379
[2]  
CRABI G, 1982, ATOM SPECTROSC, V3, P81
[3]   AN ELECTROTHERMAL SAMPLE INTRODUCTION SYSTEM FOR ICP SPECTROMETRY [J].
MATUSIEWICZ, H ;
BARNES, RM .
APPLIED SPECTROSCOPY, 1984, 38 (05) :745-747
[4]   EVALUATION OF A CONTROLLED-TEMPERATURE FURNACE ATOMIZER AS A SAMPLING DEVICE FOR INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
MATUSIEWICZ, H ;
BARNES, RM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1985, 40 :29-39
[5]   ELECTROTHERMAL VAPORIZATION FOR SAMPLE INTRODUCTION IN ATOMIC EMISSION-SPECTROMETRY [J].
NG, KC ;
CARUSO, JA .
APPLIED SPECTROSCOPY, 1985, 39 (04) :719-726
[6]   OPTIMIZATION OF ELECTROTHERMAL ATOMIZATION-INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY FOR SIMULTANEOUS MULTIELEMENT DETERMINATION [J].
SWAIDAN, HM ;
CHRISTIAN, GD .
ANALYTICAL CHEMISTRY, 1984, 56 (01) :120-122