MEASUREMENT OF BREATH AMPLITUDES - COMPARISON OF 3 NONINVASIVE RESPIRATORY MONITORS TO INTEGRATED PNEUMOTACHOGRAPH

被引:28
作者
ADAMS, JA [1 ]
ZABALETA, IA [1 ]
STROH, D [1 ]
SACKNER, MA [1 ]
机构
[1] MT SINAI MED CTR,DIV PULM DIS,MIAMI BEACH,FL 33140
关键词
RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY; RESPIGRAPH; IMPEDANCE PNEUMOGRAPHY; STRAIN GAUGES; TIDAL VOLUMES; NEWBORN;
D O I
10.1002/ppul.1950160408
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Measurement of breath amplitude (BA) and similarly tidal volume (VT) in newborn infants is the standard for detection of apnea and hypopneas. The purpose of our study was to compare the accuracy for BA by three frequently utilized noninvasive respiratory monitors: respiratory inductive plethysmography (RIP), mercury in silastic strain gauges (SG), and impedance pneumography (IP). Twenty healthy full-term infants were studied in both supine and prone postures. The RC and AB gain factors for RIP were obtained using qualitative diagnostic calibration (QDC) procedure. The electrical gain of IP was set equivalent to the BA signal of a pneumotachograph (PNT). The three devices were calibrated in the supine posture and measurements were repeated in the prone posture without changing their calibration factors. Compared to PNT, postural change did not significantly alter BA measured by RIP. The accuracy of breath-to-breath BA measurement in the prone posture was worse for IP and SG compared to RIP and PNT. In contrast to SG or IP, the accuracy of BA measurement maintained was by RIP after a postural change from supine to prone in fullterm newborns. (C) 1993 Wiley-Liss, Inc.
引用
收藏
页码:254 / 258
页数:5
相关论文
共 13 条
[1]  
Adams J A, 1993, Neonatal Intensive Care, V6, P32
[2]  
ADAMS JA, 1993, IN PRESS AM REV RESP
[3]   IMPEDANCE PNEUMOGRAPH AND MAGNETOMETER METHODS FOR MONITORING TIDAL VOLUME [J].
ASHUTOSH, K ;
GILBERT, R ;
AUCHINCLOSS, JH ;
ERLEBACHER, J ;
PEPPI, D .
JOURNAL OF APPLIED PHYSIOLOGY, 1974, 37 (06) :964-966
[4]   EFFECT OF INFANT POSITION ON BREATH AMPLITUDE MEASURED BY TRANSTHORACIC IMPEDANCE AND STRAIN-GAUGES [J].
BAIRD, TM ;
NEUMAN, MR .
PEDIATRIC PULMONOLOGY, 1991, 10 (01) :52-56
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   COMPARISON OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY AND THORACIC IMPEDANCE FOR APNEA MONITORING [J].
BROUILLETTE, RT ;
MORROW, AS ;
WEESEMAYER, DE ;
HUNT, CE .
JOURNAL OF PEDIATRICS, 1987, 111 (03) :377-383
[7]  
HELDT GP, 1982, AM REV RESPIR DIS, V138, P20
[8]   MEASUREMENT OF SEPARATE VOLUME CHANGES OF RIB CAGE AND ABDOMEN DURING BREATHING [J].
KONNO, K ;
MEAD, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1967, 22 (03) :407-&
[9]  
LI CT, 1977, J PEDIATR, V5, P223
[10]  
POLGAR G, 1965, PEDIATRICS, V36, P861