COMPARISON OF TECHNIQUES FOR DEPTH PROFILING OXYGEN IN SILICON

被引:7
作者
MEZEY, G
KOTAI, E
NAGY, T
LOHNER, L
MANUBA, A
GYULAI, J
DELINE, VR
EVANS, CA
BLATTNER, RJ
机构
[1] UNIV ILLINOIS,MAT RES LAB,URBANA,IL 61801
[2] UNIV ILLINOIS,DEPT CHEM,URBANA,IL 61801
[3] CENT RES INST PHYS,H-1525 BUDAPEST,HUNGARY
来源
NUCLEAR INSTRUMENTS & METHODS | 1979年 / 167卷 / 02期
基金
美国国家科学基金会;
关键词
D O I
10.1016/0029-554X(79)90016-8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Auger electron spectrometry (AES), Rutherford backscattering spectrometry (BS) and secondary ion mass spectrometry with Cs+ ion bombardment (Cs+/SIMS) have been experimentally compared as to their respective capabilities for depth profiling oxygen in silicon. This evaluation was conducted for depth profiling oxygen in silicon. This evaluation was conducted using standardized silicon samples with: (1) native (∼30 Å) and 80 Å thick surface oxides; (2) the same oxide layers buried beneath 1000 Å of evaporated amorphous silicon; (3) ion implanted oxygen in silicon. In general, BS was found to be the most quantitative of the three methods, particularly when using the 3.045 MeV 4He+ resonance technique for 16O in the channelling mode, but had the poorest detection limit (∼1020 atoms cm-3) and overall analytical utility. The Cs+/SIMS technique was found to have the best detection limit (5 × 1017 atoms cm-3) and general analytical applicability. Auger electron spectrometry was found to be intermediate in both regards. © 1979.
引用
收藏
页码:279 / 287
页数:9
相关论文
共 7 条
[1]   MECHANISM OF HIGH-TEMPERATURE INSTABILITY OF CUO-AG THIN-FILM SOLAR ABSORBERS [J].
BLATTNER, RJ ;
EVANS, CA ;
BRAUNDMEIER, AJ .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1977, 14 (05) :1132-1137
[2]   ELASTIC SCATTERING OF ALPHA-PARTICLES BY OXYGEN [J].
CAMERON, JR .
PHYSICAL REVIEW, 1953, 90 (05) :839-844
[3]  
MAYER JW, 1977, ION BEAM HDB MATERIA
[4]  
MEZEY G, 1976, ION BEAM SURFACE LAY, V1, P303
[5]   DEPTH PROFILE DETECTION LIMIT OF 3X10(15) ATOM CM-3 FOR AS IN SI USING CS+ BOMBARDMENT NEGATIVE SECONDARY ION MASS-SPECTROMETRY [J].
WILLIAMS, P ;
EVANS, CA .
APPLIED PHYSICS LETTERS, 1977, 30 (11) :559-561
[6]   EVALUATION OF A CESIUM PRIMARY ION-SOURCE ON AN ION MICROPROBE MASS-SPECTROMETER [J].
WILLIAMS, P ;
LEWIS, RK ;
EVANS, CA ;
HANLEY, PR .
ANALYTICAL CHEMISTRY, 1977, 49 (09) :1399-1403
[7]  
WILLIAMS PJ, COMMUNICATION