AUTOMATED SEQUENTIAL TRACE ENRICHMENT OF DIALYSATES COMBINED WITH HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND AUTOMATED HEART-CUTTING FOR THE DETERMINATION OF THE NUCLEOSIDE 1-(BETA-D-ARABINOFURANOSYL)-5-(1-PROPYNYL)URACIL AND ITS METABOLITE 5-PROPYNYLURACIL IN URINE

被引:11
作者
COOPER, JDH [1 ]
SHEUNG, CTCF [1 ]
BUICK, AR [1 ]
机构
[1] WELLCOME RES LABS,DEPT BIOANALYSIS & DRUG METAB,BECKENHAM BR3 3BS,KENT,ENGLAND
来源
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS | 1994年 / 652卷 / 01期
关键词
D O I
10.1016/0378-4347(93)E0398-A
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The use of a new configuration to control the automated sequential trace enrichment of dialysate (ASTED) system has been used to estimate 1-(beta-D-arabinofuranosyl)-5-(1-propynyl)uracil and its metabolite 5-propynyluracil in urine. The system employs ''heart-cutting'' as a means to improve the efficiency of sample preparation and reduce analysis time. Using this technique the mean within- and between-run imprecision (coefficient of variation) at three different urine analyte concentrations was found to be 1.6 and 3.6% and 1.7 and 3.3% for 5-propynyluracil and 1-(beta-D-arabinofuranosyl)-5-(1-propynyl)uracil, respectively.