OXYGEN-SATURATION OF BLOOD IN THE SUPERIOR MESENTERIC VEIN - IN-VIVO VERIFICATION OF MR-IMAGING MEASUREMENTS IN A CANINE MODEL - WORK-IN-PROGRESS

被引:36
作者
LI, KCP
WRIGHT, GA
PELC, LR
DALMAN, RL
BRITTAIN, JH
WEGMUELLER, H
LIN, DT
SONG, CK
机构
[1] STANFORD UNIV,SCH MED,DEPT SURG,STANFORD,CA 94305
[2] STANFORD UNIV,MAGNET RESONANCE SYST RES LAB,STANFORD,CA 94305
关键词
BLOOD; MR; MAGNETIC RESONANCE (MR); PULSE SEQUENCE; VASCULAR STUDIES; OXYGEN;
D O I
10.1148/radiology.194.2.7824705
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate the accuracy of magnetic resonance (MR) imaging in estimating oxygen saturation of blood (%HbO(2)) in the superior mesenteric vein (SMV) of a canine model in vivo. MATERIALS AND METHODS: MR imaging was used to measure the T2 of blood in samples obtained via a catheter placed in the SMV in seven mongrel dogs. %HbO(2) was measured with a reflectance oximeter. These measurements were obtained at the resting state, during superior mesenteric artery occlusion, and after reperfusion. MR imaging and oximeter measurements were then compared by using linear regression analysis. RESULTS: Refocusing intervals (tau(180)) of 12 and 24 msec were used for 17 and 18 %HbO(2) measurements with MR imaging, respectively. With tau(180) of 12 msec, there was an excellent correlation between MR imaging measurements and oximeter measurements (r = .969). The intercept was 5.3% and the slope was 0.959. With tau(180) Of 24 mse, r = .953, the intercept was 15.4%, and the slope was 0.817. CONCLUSION: Estimates of %HbO(2) in the SMV with MR imaging are accurate in the range of most clinical interest.
引用
收藏
页码:321 / 325
页数:5
相关论文
共 13 条
[1]   RELATIONSHIP OF BLOOD-FLOW AND OXYGEN-CONSUMPTION TO ISCHEMIC-INJURY IN THE CANINE SMALL-INTESTINE [J].
BULKLEY, GB ;
KVIETYS, PR ;
PARKS, DA ;
PERRY, MA ;
GRANGER, DN .
GASTROENTEROLOGY, 1985, 89 (04) :852-857
[2]   THE SPLANCHNIC CIRCULATION - INTRINSIC REGULATION [J].
Granger, DN ;
Kvietys, PR .
ANNUAL REVIEW OF PHYSIOLOGY, 1981, 43 :409-418
[3]  
GRANGER HJ, 1980, AM J PHYSIOL, V238, P836
[4]  
Hamar J, 1977, Adv Exp Med Biol, V94, P573
[5]  
KVIETYS PR, 1982, AM J PHYSIOL, V242, P202
[6]   MEASUREMENT OF INTRACELLULAR PH AND DEOXYHEMOGLOBIN CONCENTRATION IN DEOXYGENATED ERYTHROCYTES BY P-31 NUCLEAR MAGNETIC-RESONANCE [J].
LABOTKA, RJ .
BIOCHEMISTRY, 1984, 23 (23) :5549-5555
[7]  
LEVITT MH, 1983, ADV MAGN RESON, V11, P47
[8]   CHRONIC MESENTERIC ISCHEMIA - EVALUATION WITH PHASE-CONTRAST CINE MR-IMAGING [J].
LI, KCP ;
WHITNEY, WS ;
MCDONNELL, CH ;
FREDRICKSON, JO ;
PELC, NJ ;
DALMAN, RL ;
JEFFREY, RB .
RADIOLOGY, 1994, 190 (01) :175-179
[9]   ARTERIAL AND VENOUS-BLOOD FLOW - NONINVASIVE QUANTITATION WITH MR IMAGING [J].
PELC, LR ;
PELC, NJ ;
RAYHILL, SC ;
CASTRO, LJ ;
GLOVER, GH ;
HERFKENS, RJ ;
MILLER, DC ;
JEFFREY, RB .
RADIOLOGY, 1992, 185 (03) :809-812
[10]   ITERATIVE CARR-PURCELL TRAINS [J].
SHAKA, AJ ;
RUCKER, SP ;
PINES, A .
JOURNAL OF MAGNETIC RESONANCE, 1988, 77 (03) :606-611