COMPUTATION OF AORTIC FLOW FROM PRESSURE IN HUMANS USING A NONLINEAR, 3-ELEMENT MODEL

被引:914
作者
WESSELING, KH
JANSEN, JRC
SETTELS, JJ
SCHREUDER, JJ
机构
[1] UNIV LIMBURG,ACAD HOSP,DEPT ANESTHESIOL,6202 AZ MAASTRICHT,NETHERLANDS
[2] ERASMUS UNIV ROTTERDAM,DEPT PULM DIS,3000 DR ROTTERDAM,NETHERLANDS
关键词
AORTIC INPUT IMPEDANCE; NONLINEAR MODEL; TIME-VARYING MODEL; AORTIC ARCTANGENT PRESSURE-AREA RELATIONSHIP; AGE DEPENDENCY; SIMULATION; STROKE VOLUME; CARDIAC OUTPUT;
D O I
10.1152/jappl.1993.74.5.2566
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
We computed aortic flow pulsations from arterial pressure by simulating a nonlinear, time-varying three-element model of aortic input impedance. The model elements represent aortic characteristic impedance, arterial compliance, and systemic vascular resistance. Parameter values for the first two elements were computed from a published, age-dependent, aortic pressure-area relationship (G. J. Langewouters et al. J. Biomech. 17: 425-435, 1984). Peripheral resistance was predicted from mean pressure and model mean flow. Model flow pulsations from aortic pressure showed the visual aspects of an aortic flow curve. For evaluation we compared model mean flow from radial arterial pressure with thermodilution cardiac output estimations, 76 times, in eight open heart surgical patients. The pooled mean difference was +7%, the SD 22%. After using one comparison per patient to calibrate the model, however, we followed quantitative changes in cardiac output that occurred either during changes in the state of the patient or subsequent to vasoactive drugs. The mean deviation from thermodilution cardiac output was +2%, the SD 8%. Given these small errors the method could monitor cardiac output continuously.
引用
收藏
页码:2566 / 2573
页数:8
相关论文
共 20 条
[1]  
Broemser P, 1930, Z BIOL-MUNICH, V90, P467
[2]   ASSESSMENT OF WINDKESSEL AS A MODEL OF AORTIC INPUT IMPEDANCE [J].
BURKHOFF, D ;
ALEXANDER, J ;
SCHIPKE, J .
AMERICAN JOURNAL OF PHYSIOLOGY, 1988, 255 (04) :H742-H753
[3]  
CONWAY J, 1990, EUR HEART J, V11, P26
[4]   CONTINUOUS CARDIAC-OUTPUT MONITORING BY PULSE CONTOUR DURING CARDIAC-SURGERY [J].
JANSEN, JRC ;
WESSELING, KH ;
SETTELS, JJ ;
SCHREUDER, JJ .
EUROPEAN HEART JOURNAL, 1990, 11 :26-32
[5]   AN ADEQUATE STRATEGY FOR THE THERMODILUTION TECHNIQUE IN PATIENTS DURING MECHANICAL VENTILATION [J].
JANSEN, JRC ;
SCHREUDER, JJ ;
SETTELS, JJ ;
KLOEK, JJ ;
VERSPRILLE, A .
INTENSIVE CARE MEDICINE, 1990, 16 (07) :422-425
[6]   ESTIMATION OF STROKE VOLUME IN DOG BY A PULSE CONTOUR METHOD [J].
KOUCHOUKOS, NT ;
SHEPPARD, LC ;
MCDONALD, DA .
CIRCULATION RESEARCH, 1970, 26 (05) :611-+
[7]   THE STATIC ELASTIC PROPERTIES OF 45 HUMAN THORACIC AND 20 ABDOMINAL AORTAS INVITRO AND THE PARAMETERS OF A NEW MODEL [J].
LANGEWOUTERS, GJ ;
WESSELING, KH ;
GOEDHARD, WJA .
JOURNAL OF BIOMECHANICS, 1984, 17 (06) :425-435
[8]  
Langewouters GJ, 1985, BIOMECHANICS CURRENT
[9]   LEFT VENTRICULAR OUTPUT DERIVED FROM TIME-DERIVATIVE AND PHASE VELOCITIES OF AORTIC PRESSURE WAVE [J].
MCDONALD, DA ;
NICHOLS, WW .
MEDICAL & BIOLOGICAL ENGINEERING, 1973, 11 (06) :678-690
[10]  
MCDONALD DA, 1974, BLOOD FLOW ARTERIES