EVANESCENT-WAVE IMMUNOPROBE WITH HIGH BIVALENT ANTIBODY-ACTIVITY

被引:11
作者
FELDMAN, SF
UZGIRIS, EE
PENNEY, CM
GUI, JY
SHU, EY
STOKES, EB
机构
[1] General Electric Corporate Research and Development, Schenectady
关键词
EVANESCENT WAVE FIBER IMMUNOSENSOR; ANTIBODY ANTIGEN KINETICS; UNSTIRRED DEAD LAYER THICKNESS; IMMUNOSENSOR RESPONSE; OPTICAL FIBER SENSOR;
D O I
10.1016/0956-5663(95)96889-7
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
We have determined the kinetic response of an evanescent wave optical fibre immunosensor and its absolute sensitivity. Using both kinetic methods and optical determinations of bound antigen, we infer that there are 2.4 x 10(11) active antibodies per cm(2) probe area. We estimate that 75% of the active antibodies are in the bivalent form, with both binding sites capable of binding antigen. Using the kinetic data and optical measurements of the number of antigens bound to the fibre, we estimate that the unstirred dead layer thickness next to the fibre surface in the stirred solution, across which the antigens must diffuse, is 55 mu m. Approximately 10(9) FITC labelled antigens are necessary to achieve a threshold signal, defined as 1% of the signal obtained when the fibre is saturated with antigen.
引用
收藏
页码:423 / 434
页数:12
相关论文
共 23 条
[21]  
Sutherland, Dahne, Place, Ringnose, Optical detection of antibody-antigen reactions at a glass-liquid interface, Clin. Chem., 30, pp. 1533-1538, (1984)
[22]  
Thompson, Ligler, Chemistry and technology of evanescent wave biosensors, Biosensors with Fiberoptics, pp. 111-138, (1991)
[23]  
Thompson, Villarruel, Waveguide-binding sensor for use with assays, US Patent #5,061,857, (1991)