ACCURACY OF INTERFACE PRESSURE MEASUREMENT SYSTEMS

被引:35
作者
ALLEN, V [1 ]
RYAN, DW [1 ]
LOMAX, N [1 ]
MURRAY, A [1 ]
机构
[1] FREEMAN RD HOSP,INTENSIVE THERAPY UNIT,NEWCASTLE TYNE NE7 7DN,TYNE & WEAR,ENGLAND
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1993年 / 15卷 / 04期
关键词
TRANSDUCERS; INTERFACE PRESSURE; MEASUREMENT ERRORS;
D O I
10.1016/0141-5425(93)90014-P
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Interface pressure measurement is needed to assess beds designed to prevent pressure sores, so it is therefore important to establish the accuracy of interface pressure measuring systems. In this study, the Talley SA500 pressure evaluator (with 28 mm and 100 mm sensor pads), the DIPE (with 100 mm sensor pad), and a water-filled bladder system (with 0.1 ml and 0.3 ml water) were assessed. Measurement errors were evaluated using a loading system with pressures up to 7.4 kPa (55 mm Hg) in steps of 0.9 kPa (6.9 mm Hg). All systems tested over-measured interface pressure, the error being approximately linearly proportional to the loading pressure. The repeatability for a given system was approximately constant. The mean error (+/- SD) (%) and repeatability (kPa) for the systems were: 28 mm Talley 12 +/- 1%, +/- 0.07 kPa; 100 mm Talley 15 +/- 1%, +/-0.07 kPa; DIPE 27 +/- 3%, +/- 0.12 kPa; 0.1 ml water bladder 17 +/- 1%, +/- 0.13 kPa; 0.3 ml water bladder 26 +/- 3%, +/- 0.07 kPa. Different interfaces affected accuracy markedly, and repeatability was affected when an inhomogeneous interface was used. The study shows that the errors associated with interface pressure measurement systems can be substantial, and can vary from one system to another.
引用
收藏
页码:344 / 348
页数:5
相关论文
共 24 条
[1]   CHANGES IN TRANS-CUTANEOUS OXYGEN-TENSION AS A RESULT OF PROLONGED PRESSURES AT THE SACRUM [J].
BADER, DL ;
GANT, CA .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1988, 9 (01) :33-40
[2]   PRESSURE DISTRIBUTION UNDER THE ISCHIUM OF NORMAL SUBJECTS [J].
BADER, DL ;
HAWKEN, MB .
JOURNAL OF BIOMEDICAL ENGINEERING, 1986, 8 (04) :353-357
[3]   DEVICE FOR MEASURING SOFT-TISSUE INTERFACE PRESSURES [J].
BARBENEL, JC ;
SOCKALINGHAM, S .
JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (06) :519-522
[4]  
BEEBE D, 1991, PREVENTION PRESSURE, P155
[5]  
BENNETT L, 1979, ARCH PHYS MED REHAB, V60, P309
[6]   PREVENTING PRESSURE SORES IN HOSPITAL - CONTROLLED TRIAL OF A LARGE-CELLED RIPPLE MATTRESS [J].
BLISS, MR ;
MCLAREN, R ;
EXTONSMI.AN .
BRITISH MEDICAL JOURNAL, 1967, 1 (5537) :394-&
[7]   A CLINICAL-EVALUATION OF THE AIR-FLUIDIZED BED IN A GENERAL PLASTIC-SURGERY UNIT [J].
BOORMAN, JG ;
CARR, S ;
KEMBLE, JVH .
BRITISH JOURNAL OF PLASTIC SURGERY, 1981, 34 (02) :165-168
[8]  
CLARK M, 1987, CARE SC PRACT JUN, P5
[9]  
DINSDALE SM, 1974, ARCH PHYS MED REHAB, V55, P147
[10]  
Ferguson-Pell M. W., 1980, Engineering in Medicine, V9, P209, DOI 10.1243/EMED_JOUR_1980_009_052_02