AUTOMATION AND VALIDATION OF THE HIGH-PERFORMANCE LIQUID CHROMATOGRAPHIC-RADIOIMMUNOASSAY METHOD FOR THE DETERMINATION OF LACIDIPINE IN PLASMA

被引:19
作者
BRAGGIO, S
SARTORI, S
ANGERI, F
PELLEGATTI, M
机构
[1] Research Laboratories, Glaxo S.p.A., 37100 Verona
来源
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS | 1995年 / 669卷 / 02期
关键词
D O I
10.1016/0378-4347(95)00128-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The automation and validation of the HPLC-radioimmunoassay (RIA) method for the determination of lacidipine are reported. The solid-phase extraction step was automated by the introduction of the ASPEC system. A two-column system was adopted for the HPLC purification. The RIA was converted from heterogeneous to homogeneous by the scintillation proximity assay system and automated using an automatic dilution system. All characteristics in terms of accuracy, precision, specificity, and linearity resulted similar to the manual version. The quantification limit was set to 40 pg/ml. The new version of the method increased the number of samples assayed per month two- to three-fold.
引用
收藏
页码:383 / 389
页数:7
相关论文
共 3 条
[1]  
HILDEBRAND M, 1992, EICOSANOIDS, V5, P5
[2]   VALIDATION OF A HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC RADIOIMMUNOASSAY METHOD FOR THE DETERMINATION OF LACIDIPINE IN PLASMA [J].
PELLEGATTI, M ;
BRAGGIO, S ;
SARTORI, S ;
FRANCESCHETTI, F ;
BOLELLI, GF .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1992, 573 (01) :105-111
[3]   SCINTILLATION PROXIMITY ASSAY - A SENSITIVE AND CONTINUOUS ISOTOPIC METHOD FOR MONITORING LIGAND RECEPTOR AND ANTIGEN-ANTIBODY INTERACTIONS [J].
UDENFRIEND, S ;
GERBER, L ;
NELSON, N .
ANALYTICAL BIOCHEMISTRY, 1987, 161 (02) :494-500