EXPERIMENTAL-VERIFICATION OF A MODEL FOR PREDICTING TRANSIENT TEMPERATURE DISTRIBUTIONS BY FOCUSED ULTRASOUND

被引:3
作者
CHIN, RB
ZAGZEBSKI, JA
MADSEN, EL
机构
[1] UNIV WISCONSIN,DEPT MED PHYS,MADISON,WI 53706
[2] UNIV WISCONSIN,DEPT HUMAN ONCOL,MADISON,WI 53706
[3] UNIV WISCONSIN,DEPT RADIOL,MADISON,WI 53706
关键词
D O I
10.1088/0031-9155/36/9/001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A thermal model for predicting time-dependent temperature distributions during ultrasound heating was tested quantitatively. The relevant thermal processes incorporated in the model are heat conduction and ultrasound power absorption, and the required input parameters include the absolute ultrasound power, the shape and frequency of the ultrasound transducer, and the thermal and acoustical properties of the medium. Testing was done by heating an ultrasonically tissue-mimicking phantom with a 525 kHz, single element, focused source. The phantom has muscle-like acoustical properties and contains an array of copper-constantan thermocouples. Time-dependent temperature changes predicted with the thermal model were in very good agreement with those measured in the phantom, verifying the validity of the model for use in optimizing an ultrasound applicator design for a specific treatment situation.
引用
收藏
页码:1153 / 1164
页数:12
相关论文
共 27 条
[21]  
2-7
[22]   MEASUREMENT OF ULTRASOUND ENERGY DENSITY DISTRIBUTIONS INVIVO [J].
MARTIN, CJ ;
HYNYNEN, K ;
WATMOUGH, DJ .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1984, 10 (06) :701-708
[23]   SIMULATIONS OF SCANNED FOCUSED ULTRASOUND HYPERTHERMIA - THE EFFECTS OF SCANNING SPEED AND PATTERN ON THE TEMPERATURE-FLUCTUATIONS AT THE FOCAL DEPTH [J].
MOROS, EG ;
ROEMER, RB ;
HYNYNEN, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (05) :552-560
[24]  
NOYE J, 1982, NUMERICAL SOLUTION P
[25]   THEORY OF FOCUSING RADIATORS [J].
ONEIL, HT .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1949, 21 (05) :516-526
[26]   SIMULATION OF FOCUSED, SCANNED ULTRASONIC HEATING OF DEEP-SEATED TUMORS - THE EFFECT OF BLOOD PERFUSION [J].
ROEMER, RB ;
SWINDELL, W ;
CLEGG, ST ;
KRESS, RL .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1984, 31 (05) :457-466
[27]   A SURVEY OF COMPUTER-SIMULATIONS OF HYPERTHERMIA TREATMENTS [J].
STROHBEHN, JW ;
ROEMER, RB .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (01) :136-149