Near-infrared spectroscopy versos compartment pressure for the diagnosis of lower extremity compartmental syndrome using electromyography-determined measurements of neuromuscular function

被引:69
作者
Gentilello, LM
Sanzone, A
Wang, LL
Liu, PY
Robinson, L
机构
[1] Univ Washington, Harborview Med Ctr, Dept Surg, Sch Med, Seattle, WA 98104 USA
[2] Univ Washington, Harborview Med Ctr, Dept Orthoped, Sch Med, Seattle, WA 98104 USA
[3] Univ Washington, Harborview Med Ctr, Dept Rehabil Med, Sch Med, Seattle, WA 98104 USA
来源
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE | 2001年 / 51卷 / 01期
关键词
compartmental syndrome; near-infrared spectroscopy; fractures; oximetry; musculoskeletal physiology; ischemia; orthopedic equipment; orthopedics;
D O I
10.1097/00005373-200107000-00001
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Compartmental syndrome (CS) is difficult to diagnose in intensive care unit patients. Compartment perfusion pressure (CPP) is an invasive, indirect measure of ischemia, Near-infrared spectroscopy is noninvasive, and directly measures ischemia by transmitting light through tissues at wavelengths that react with hemoglobin to provide percent tissue oxygen saturation (Sto(2)), Animal studies demonstrate that Sto(2) is superior to CPP for detecting CS, However, there are no studies in humans comparing Sto(2) with CPP, We hypothesized that Sto(2) can reliably detect CS, and is superior to CPP, Methods: CS was induced in 15 human volunteers using a standard calf compression model. At 30-minute intervals, compression was increased to reduce Sto(2) from baseline (86% +/- 4%) to 60%, 40%, 20%, and < 10%, with simultaneous recording of CPP, Outcome variables included deep peroneal nerve conduction assessed by electromyography, cutaneous peroneal nerve sensitivity using Semmes-Weinstein monofilaments, and pain (visual analog scale). Results: Both Sto(2) and CPP significantly correlated with all ischemia outcome variables (p < 0.001), Receiver operating characteristic curves of deep peroneal nerve conduction demonstrated that Sto(2) had higher sensitivity than CPP for detecting > 50% block. For example, when specificity was 83% for Sto(2) and 84% for CPP, sensitivity was 85% versus 56%, respectively (p = 0.02), When specificity for both was 72%, sensitivity was 94% for Sto(2) versus 76% for CPP (p = 0.04). Conslusion: In intensive care unit patients who cannot alert physicians to symptoms, near-infrared spectroscopy may help clinicians to avoid delayed or unnecessary prophylactic fasciotomy, and provides the benefits of a continuous, noninvasive monitoring technique.
引用
收藏
页码:1 / 8
页数:8
相关论文
共 28 条
  • [1] Near-infrared spectroscopy: A potential method for continuous, transcutaneous monitoring for compartmental syndrome in critically injured patients
    Arbabi, S
    Brundage, SI
    Gentilello, LM
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1999, 47 (05): : 829 - 833
  • [2] THE REPEATABILITY OF TESTING WITH SEMMES-WEINSTEIN MONOFILAMENTS
    BELLKROTOSKI, J
    TOMANCIK, E
    [J]. JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1987, 12A (01): : 155 - 161
  • [3] Fasciotomy, chronic venous insufficiency, and the calf muscle pump
    Bermudez, K
    Knudson, MM
    Morabito, D
    Kessel, O
    [J]. ARCHIVES OF SURGERY, 1998, 133 (12) : 1356 - 1361
  • [4] The effect of antecedent ischemia on the tolerance of skeletal muscle to increased interstitial pressure
    Bernot, M
    Gupta, R
    Dobrasz, J
    Chance, B
    Heppenstall, RB
    Sapega, A
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 1996, 10 (08) : 555 - 559
  • [5] ASSESSMENT AND MANAGEMENT OF FOOT DISEASE IN PATIENTS WITH DIABETES
    CAPUTO, GM
    CAVANAGH, PR
    ULBRECHT, JS
    GIBBONS, GW
    KARCHMER, AW
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (13) : 854 - 860
  • [6] EFFECT OF ISCHEMIA ON SENSORY POTENTIALS OF NORMAL SUBJECTS OF DIFFERENT AGES
    CARUSO, G
    LABIANCA, O
    FERRANNINI, E
    [J]. JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1973, 36 (03) : 455 - 466
  • [7] TISSUE PRESSURE AND PERFUSION IN COMPARTMENT SYNDROME
    CLAYTON, JM
    HAYES, AC
    BARNES, RW
    [J]. JOURNAL OF SURGICAL RESEARCH, 1977, 22 (04) : 333 - 339
  • [8] DAHN I, 1967, CLIN SCI, V32, P467
  • [9] Monitoring for compartmental syndrome using near-infrared spectroscopy: A noninvasive, continuous, transcutaneous monitoring technique
    Garr, JL
    Gentilello, LM
    Cole, PA
    Mock, CN
    Matsen, FA
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1999, 46 (04): : 613 - 616
  • [10] SENSIBILITY TESTING IN PERIPHERAL-NERVE COMPRESSION SYNDROMES - AN EXPERIMENTAL-STUDY IN HUMANS
    GELBERMAN, RH
    SZABO, RM
    WILLIAMSON, RV
    DIMICK, MP
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1983, 65 (05) : 632 - 638