High-resolution fluorescent particle-tracking flow visualization within an intraventricular axial flow left ventricular assist device

被引:11
作者
Kerrigan, JP
Yamazaki, K
Meyer, RK
Mori, T
Otake, Y
Outa, E
Umezu, M
Borovetz, HS
Kormos, RL
Griffith, BP
Koyanagi, H
Antaki, JF
机构
[1] UNIV PITTSBURGH,MED CTR,ARTIFICIAL HEART & LUNG PROGRAM,PITTSBURGH,PA 15219
[2] TOKYO WOMENS MED COLL,TOKYO 162,JAPAN
[3] SUN MED TECHNOL RES CORP,NAGANO,JAPAN
[4] WASEDA UNIV,TOKYO,JAPAN
关键词
flow visualization; artificial organs; fluorescent image tracking velocimetry; axial pump;
D O I
10.1111/j.1525-1594.1996.tb04476.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Flow visualization is typically applied in blood pump development to both confirm the design expectations and identify regions that may be predisposed to blood element deposition and trauma. Rotary pumps, in particular, place high demands on the technique chosen to visualize the flow given the limited visual accessibility of the flow path and the high impeller speeds. Fluorescent image-tracking velocimetry currently is used at the University of Pittsburgh Medical Center to visualize flow accurately inside of these pumps both qualitatively and quantitatively. Flow patterns under steady conditions within an intraventricular axial flow, left ventricular assist pump (prototype No. 7, SUN Medical Technology Research Corporation, Nagano, Japan) were investigated using this technique. The flow fields at the impeller-stator interface and at the pump outlet were given specific attention. This allowed the assessment of the fluid dynamics throughout the hydrodynamic design limits of the pump.
引用
收藏
页码:534 / 540
页数:7
相关论文
共 11 条
  • [1] AHMED SN, 1987, FLUID MECH
  • [2] COMPUTATIONAL FLOW OPTIMIZATION OF ROTARY BLOOD PUMP COMPONENTS
    ANTAKI, JF
    GHATTAS, O
    BURGREEN, GW
    HE, BC
    [J]. ARTIFICIAL ORGANS, 1995, 19 (07) : 608 - 615
  • [3] 3-DIMENSIONAL NUMERICAL PREDICTION OF STRESS LOADING OF BLOOD PARTICLES IN A CENTRIFUGAL PUMP
    BLUDSZUWEIT, C
    [J]. ARTIFICIAL ORGANS, 1995, 19 (07) : 590 - 596
  • [4] Kerrigan J P, 1993, ASAIO J, V39, pM639
  • [5] KERRIGAN JP, 1994, P ASME FLUIDS ENG C, V191, P155
  • [6] VISUALIZATION STUDY OF FLOW IN AXIAL-FLOW INDUCER
    LAKSHMINARAYANA, B
    [J]. JOURNAL OF BASIC ENGINEERING, 1972, 94 (04): : 777 - 787
  • [7] LEONARD EF, 1993, CARDIOVASC PATHOL, V2, pS3
  • [8] PAPANTONIS D, 1991, 1991 P INT WORKSH RO, P63
  • [9] SHAFFER F, 1992, P ASME FLUIDS ENG C
  • [10] EVALUATION OF BIOPROSTHETIC VALVE-ASSOCIATED THROMBUS IN VENTRICULAR ASSIST DEVICE PATIENTS
    WAGNER, WR
    JOHNSON, PC
    KORMOS, RL
    GRIFFITH, BP
    [J]. CIRCULATION, 1993, 88 (05) : 2023 - 2029