The use of a thin thiol-functionalized silica film modified glassy carbon electrode in the determination of Hg(II) ions in a natural water sample is described. A typical measurement involves two successive steps: a glassy carbon electrode coated with a thin mesoporous silica film containing 10% of mercaptopropyl groups, according to the MPTMS/TEOS ratio in the starting sol-gel, was first immersed into the accumulation medium for 15 min, then removed, and finally transferred into a detection solution containing KCl 1.0 mol L-1 where detection was performed by anodic stripping voltammetry. In this medium the previously accumulated Hg2+ species complexed by the thiol groups in an open circuit preconcentration step is then directly reduced at -0.6 V during 60 s prior to be quantified by a differential pulse anodic scan from -0.6 to 0.3 V (vs. Ag/AgCl). A stripping peak appeared at about -0.01 V, which is directly proportional to the quantity of the analyte previously accumulated into the film. The best results were obtained under the following conditions: 100 mV pulse amplitude and 10 mV s(-1) scan rate in 1.0 mol L-1 KCl solution pH 2.0. Using such parameters a linear dynamic range from 1.00 to 10.0 x 10(-8) mol L-1 Hg(II) was observed with a limit of detection of 4.3 nmol L-1 for an accumulation time of 15 min. Hg(II) spiked in a natural water sample was determined between 97.0 and 101.4% mean recovery at 10(-8) mol L-1 level. The results indicate that this electrode is sensitive and selective for the Hg(II)determination.
引用
收藏
页码:2301 / 2309
页数:9
相关论文
共 42 条
[1]   CHEMICALLY MODIFIED ELECTRODE FOR THE SIMULTANEOUS DETERMINATION OF TRACE-METALS AND SPECIATION ANALYSIS [J].
AGRAZ, R ;
SEVILLA, MT ;
HERNANDEZ, L .
ANALYTICA CHIMICA ACTA, 1993, 273 (1-2) :205-212
[2]   VOLTAMMETRIC QUANTIFICATION AND SPECIATION OF MERCURY-COMPOUNDS [J].
AGRAZ, R ;
SEVILLA, MT ;
HERNANDEZ, L .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1995, 390 (1-2) :47-57
[3]   Biomimetic pathways for assembling inorganic thin films [J].
Aksay, IA ;
Trau, M ;
Manne, S ;
Honma, I ;
Yao, N ;
Zhou, L ;
Fenter, P ;
Eisenberger, PM ;
Gruner, SM .
SCIENCE, 1996, 273 (5277) :892-898
[4]   Field sampling, preconcentration and determination of mercury species in river waters [J].
Blanco, RM ;
Villanueva, MT ;
Uría, JES ;
Sanz-Medel, A .
ANALYTICA CHIMICA ACTA, 2000, 419 (02) :137-144
[5]   Determination of methylmercury in natural water samples by steam distillation and gas chromatography atomic fluorescence spectrometry [J].
Bowles, KC ;
Apte, SC .
ANALYTICAL CHEMISTRY, 1998, 70 (02) :395-399
[6]  
Brinker CJ, 1999, ADV MATER, V11, P579, DOI 10.1002/(SICI)1521-4095(199905)11:7<579::AID-ADMA579>3.3.CO
[7]  
2-I
[8]   A general one-pot process leading to highly functionalised ordered mesoporous silica films [J].
Cagnol, F ;
Grosso, D ;
Sanchez, C .
CHEMICAL COMMUNICATIONS, 2004, (15) :1742-1743
[9]  
CARRAMOLINO GC, 2008, ANAL CHIM ACTA, V614, P103
[10]   Evaluation of a carbon paste electrode modified with organofunctionalised SBA-15 nanostructured silica in the simultaneous determination of divalent lead, copper and mercury ions [J].
Cesarino, Ivana ;
Marino, Glimaldo ;
do Rosario Matos, Jivaldo ;
Gomes Cavalheiro, Eder Tadeu .
TALANTA, 2008, 75 (01) :15-21