Application of slurry sample introduction to a low-power inductively coupled plasma atomic emission spectrometer incorporating a mini-torch

被引:16
作者
Hyun, JH
Lim, HB
Lim, CH
机构
[1] DANKOOK UNIV,DEPT CHEM,YOUNGSAN KU,SEOUL 140714,SOUTH KOREA
[2] KOREA TESTING LAB IND TECHNOL,MAT ANAL TEAM,KURO KU,SEOUL 152053,SOUTH KOREA
关键词
Al2O3; slurry nebulization; inductively coupled plasma spectrometry; atomic emission spectrometry; low power; mini-torch; MASS-SPECTROMETRY; PARTICLE-SIZE; NEBULIZATION; ATOMIZATION; SUSPENSIONS; POWDERS; SOLIDS; COAL;
D O I
10.1016/S0003-2670(96)00579-X
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学];
摘要
The use of a low-power ICP (0.65 kW) and a mini-torch (outer tube: 12.6 mm o.d.) for slurry sampling of Al2O3 powders is described. In order to find the optimum conditions for mini-torch ICP-AES, the influence of particle size and aerosol gas flow on the signal intensities was studied. The emission intensities obtained with slurries increased as the particle size decreased and maximized in aqueous solution. The intensities obtained with slurries also increased with aerosol flow rate, but were still smaller than those from solution. The intensity difference between particles of size 0.2-30 mu m and solutions increased with increased aerosol flow rate up to a maximum at 0.81 min(-1). Dependence of the emission intensity and its relative standard deviation on the pH of the slurries in the mini-torch ICP was also studied. The best signal-to-noise ratio was obtained at pH 3, when the pH of the suspension was varied in the range 1-5, The analysis of JCRM 032 (Japan Ceramic Reference Materials) Al2O3 powders for trace impurities using slurry nebulization under the optimized conditions provided a recovery of > 85% for MgO and Fe2O3 and 67% for Na2O and SiO2, with precision of < 4.3% RSD, The analytical results were also compared with those estimated for the same material, using decomposition method in sealed PTFE vessels at high pressure.
引用
收藏
页码:83 / 89
页数:7
相关论文
共 19 条
[1]
DIRECT ATOMIC SPECTROMETRIC ANALYSIS BY SLURRY ATOMIZATION .8. FLOW-INJECTION INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
AMBROSE, AJ ;
EBDON, L ;
FOULKES, ME ;
JONES, P .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (02) :219-222
[2]
NONREFRACTORY SLURRIES IN THE ICP - SIZE CONTROL AND RECOVERY [J].
CHEN, C ;
MCCREARY, TW .
APPLIED SPECTROSCOPY, 1994, 48 (03) :410-412
[3]
DIRECT ATOMIC SPECTROMETRIC ANALYSIS BY SLURRY ATOMIZATION .7. ANALYSIS OF COAL USING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EBDON, L ;
FOULKES, ME ;
PARRY, HGM ;
TYE, CT .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (06) :753-761
[4]
[5]
OPTIMIZED SIMULTANEOUS MULTIELEMENT ANALYSIS OF ENVIRONMENTAL SLURRY SAMPLES BY INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY USING A SEGMENTED ARRAY CHARGE-COUPLED-DEVICE DETECTOR [J].
EBDON, L ;
FOULKES, M ;
OHANLON, K .
ANALYTICA CHIMICA ACTA, 1995, 311 (02) :123-134
[6]
PARTICLE-SIZE EFFECTS ON KAOLIN SLURRY ANALYSIS BY INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
EBDON, L ;
COLLIER, AR .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1988, 43 (4-5) :355-369
[7]
EFFECT OF COLLOIDAL STABILITY OF CERAMIC SUSPENSIONS ON NEBULIZATION OF SLURRIES FOR INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
FARINAS, JC ;
MORENO, R ;
MERMET, JM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (08) :841-849
[8]
FRY RC, 1988, APPL SPECTROSC, V42, P381
[9]
SLURRY NEBULIZATION INDUCTIVELY-COUPLED PLASMA SPECTROMETRY - THE FUNDAMENTAL PARAMETERS DISCUSSED [J].
GOODALL, P ;
FOULKES, ME ;
EBDON, L .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (13) :1563-1577
[10]
DIRECT ANALYSIS OF SOLIDS BY ULTRASONIC SLURRY ELECTROTHERMAL VAPORIZATION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
GREGOIRE, DC ;
MILLERIHLI, NJ ;
STURGEON, RE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (05) :605-610