Log-file analysis of accuracy of beam localization for brain tumor treatment by CyberKnife

被引:7
作者
Okamoto, Hiroyuki [1 ]
Hamada, Minoru [1 ]
Sakamoto, Emi [1 ]
Wakita, Akihisa [1 ]
Nakamura, Satoshi [1 ]
Kato, Toru [1 ]
Abe, Yoshihisa [1 ]
Takahashi, Kana [1 ]
Igaki, Hiroshi [1 ]
Itami, Jun [1 ]
机构
[1] Natl Canc Ctr, Dept Radiat Oncol, Tokyo 1040045, Japan
关键词
ROBOTIC RADIOSURGERY; TRACKING SYSTEM; PATIENT MOTION; QA;
D O I
10.1016/j.prro.2016.01.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 [肿瘤学];
摘要
Purpose: The CyberKnife system generates log-files including actual treatment parameters for each procedure. In this study, log-files were analyzed to evaluate the mechanical uncertainty in beam localization over the long term (approximately 1 year), as were patterns of patient movements during brain tumor treatments using CyberKnife. Methods and materials: The clinical to planning target volume (CTV-PTV) margin in clinical use was examined based on this analysis. Log-file analysis was performed using data from 140 brain tumor patients (267 treatment plans; 27,166 beams; approximately 66 beams/fraction), who underwent CyberKnife stereotactic radiosurgery and radiation therapy. We calculated a mean error and 2 standard deviations (2 sigma) for this population. Additionally, we calculated the radius R-95% spatially covering 95% of all error vectors. Results: The mean mechanical uncertainties of CyberKnife brain tumor treatment were found to be 0.07, 0.01, and -0.09 mm in the +inferior/-superior, +left/-right, and +anterior/-posterior directions, respectively. The mean (2 sigma) of R-95% was 1.02 (0.42) mm. A smaller degree of correlation between patient movement and R-95% was observed. Conclusion: The CyberKnife is robust in tracking accuracy, regardless of patient movement. The effectiveness of log-file analysis was demonstrated regarding quality control for monitoring beam localization in the CyberKnife system. The CTV-PTV margin of 2.0 mm was found to be adequate in brain tumor treatments using the CyberKnife. (C) 2016 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:E361 / E367
页数:7
相关论文
共 15 条
[1]
Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries [J].
Agnew, A. ;
Agnew, C. E. ;
Grattan, M. W. D. ;
Hounsell, A. R. ;
McGarry, C. K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (09) :N49-N63
[2]
Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT [J].
Agnew, Christina E. ;
Irvine, Denise M. ;
McGarry, Conor K. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (06) :204-216
[3]
A Varian DynaLog file-based procedure for patient dose-volume histogram-based IMRT QA [J].
Calvo-Ortega, Juan F. ;
Teke, Tony ;
Moragues, Sandra ;
Pozo, Miquel ;
Casals, Joan .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02) :100-109
[4]
Report of AAPM TG 135: Quality assurance for robotic radiosurgery [J].
Dieterich, Sonja ;
Cavedon, Carlo ;
Chuang, Cynthia F. ;
Cohen, Alan B. ;
Garrett, Jeffrey A. ;
Lee, Charles L. ;
Lowenstein, Jessica R. ;
d'Souza, Maximian F. ;
Taylor, David D., Jr. ;
Wu, Xiaodong ;
Yu, Cheng .
MEDICAL PHYSICS, 2011, 38 (06) :2914-2936
[5]
The design, physical properties and clinical utility of an iris collimator for robotic radiosurgery [J].
Echner, G. G. ;
Kilby, W. ;
Lee, M. ;
Earnst, E. ;
Sayeh, S. ;
Schlaefer, A. ;
Rhein, B. ;
Dooley, J. R. ;
Lang, C. ;
Blanck, O. ;
Lessard, E. ;
Maurer, C. R., Jr. ;
Schlegel, W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (18) :5359-5380
[6]
PATIENT MOTION AND TARGETING ACCURACY IN ROBOTIC SPINAL RADIOSURGERY: 260 SINGLE-FRACTION FIDUCIAL-FREE CASES [J].
Fuerweger, Christoph ;
Drexler, Christian ;
Kufeld, Markus ;
Muacevic, Alexander ;
Wowra, Berndt ;
Schlaefer, Alexander .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03) :937-945
[7]
CLINICAL ACCURACY OF THE RESPIRATORY TUMOR TRACKING SYSTEM OF THE CYBERKNIFE: ASSESSMENT BY ANALYSIS OF LOG FILES [J].
Hoogeman, Mischa ;
Prevost, Jean-Briac ;
Nuyttens, Joost ;
Poell, Johan ;
Levendag, Peter ;
Heijmen, Ben .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (01) :297-303
[8]
Effect of residual patient motion on dose distribution during image-guided robotic radiosurgery for skull tracking based on log file analysis [J].
Inoue, Mitsuhiro ;
Shiomi, Hiroya ;
Sato, Kengo ;
Taguchi, Junichi ;
Okawa, Kohei ;
Inada, Kosaku ;
Murai, Taro ;
Koike, Izumi ;
Tatewaki, Koshi ;
Ota, Seiji ;
Inoue, Tomio .
JAPANESE JOURNAL OF RADIOLOGY, 2014, 32 (08) :461-466
[9]
Validation of dynamic MLC-controller log files using a two-dimensional diode array [J].
Li, JG ;
Dempsey, JF ;
Ding, L ;
Liu, CR ;
Palta, JR .
MEDICAL PHYSICS, 2003, 30 (05) :799-805
[10]
Litzenberg Dale W, 2002, J Appl Clin Med Phys, V3, P63, DOI 10.1120/1.1449362