Retrieval of the Gunther tulip and OptEase filters using the Recovery Cone Removal System

被引:7
作者
Miller, Michael J.
Smith, Tony P.
Stavas, Joseph M.
机构
[1] Duke Univ, Med Ctr, Div Intervent Radiol, Dept Radiol, Durham, NC 27710 USA
[2] Univ N Carolina, Sch Med, Dept Radiol, Chapel Hill, NC 27599 USA
关键词
D O I
10.1097/01.RVI.0000217952.62724.E7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Recovery Cone Removal System was evaluated in vitro to determine its feasibility as an alterative retrieval device for the Gunther tulip and OptEase inferior vena cava filters. The cone removal system was then used electively in eight patients to determine feasibility and safety in vivo and subsequently in a single patient when wire-snare removal failed. The Recovery Cone Removal System allowed in vitro and in vivo retrieval in all cases without difficulty or complication. The Recovery Cone Removal System offers an alternative to the wire-snare-based retrieval systems for removal of the Gunther tulip and OptEase filters.
引用
收藏
页码:899 / 902
页数:4
相关论文
共 12 条
[1]   Initial experience in humans with a new retrievable inferior vena cava filter [J].
Asch, MR .
RADIOLOGY, 2002, 225 (03) :835-844
[2]   The Gunther Tulip retrievable filter:: Prolonged temporary filtration by repositioning within the inferior vena cava [J].
de Gregorio, MA ;
Gamboa, P ;
Gimeno, MJ ;
Madariaga, B ;
Tobío, R ;
Herrera, M ;
Medrano, J ;
Mainar, A ;
Alfonso, R .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (10) :1259-1265
[3]   Animal experience in the Gunther Tulip retrievable inferior vena cava filter [J].
de Gregorio, MA ;
Gimeno, MJ ;
Tobio, R ;
Lostalé, F ;
Mainar, A ;
Beltrán, JM ;
Madariaga, B ;
Alfonso, ER ;
Medrano, J ;
Viloria, A .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 24 (06) :413-417
[4]   A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis [J].
Decousus, H ;
Leizorovicz, A ;
Parent, F ;
Page, Y ;
Tardy, B ;
Girard, P ;
Laporte, S ;
Faivre, R ;
Charbonnier, B ;
Barral, FG ;
Huet, Y ;
Simonneau, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 338 (07) :409-415
[5]   Experience with the recovery filter as a retrievable inferior vena cava filter [J].
Grande, WJ ;
Trerotola, SO ;
Reilly, PM ;
Clark, TWI ;
Soulen, MC ;
Patel, A ;
Shlansky-Goldberg, RD ;
Tuite, CM ;
Solomon, JA ;
Mondschein, JI ;
Fitzpatrick, MK ;
Stavropoulos, SW .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 16 (09) :1189-1193
[6]   Difficult retrieval of a recovery IVC filter [J].
Hagspiel, KD ;
Leung, DA ;
Aladdin, M ;
Spinosa, DJ ;
Matsumoto, AH ;
Angle, JF .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 15 (06) :645-647
[7]   Gunther tulip retrievable vena cava filter: Results from the registry of the Canadian Interventional Radiology Association [J].
Millward, SF ;
Oliva, VL ;
Bell, SD ;
Valenti, DA ;
Rasuli, P ;
Asch, M ;
Hadziomerovic, A ;
Kachura, JR .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (09) :1053-1058
[8]   Percutaneous retrieval of the Tulip vena cava filter:: Feasibility, short- and long-term changes -: An experimental study in dogs [J].
Neuerburg, JM ;
Handt, S ;
Beckert, K ;
Tonn, K ;
Rasmussen, E ;
Hunter, D ;
Günther, RW .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 24 (06) :418-423
[9]   Evaluation of the retrievability of the OptEase IVC filter in an animal model [J].
Reekers, JA ;
Hoogeveen, YL ;
Wijnands, M ;
Bosma, G ;
Mulder, R ;
Oliva, VL .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 15 (03) :261-267
[10]   Role of prophylactic temporary inferior vena cava filters placed at the ICU bedside under intravascular ultrasound guidance in patients with multiple trauma [J].
Rosenthal, D ;
Wellons, ED ;
Levitt, AB ;
Shuler, FW ;
O'Conner, RE ;
Henderson, VJ .
JOURNAL OF VASCULAR SURGERY, 2004, 40 (05) :958-963