Use of immobilized amyloglucosidase as chiral selector in chromatography.: Control of enantioselective retention and resolution in liquid chromatography

被引:19
作者
Strandberg, A [1 ]
Nyström, A [1 ]
Behr, S [1 ]
Karlsson, A [1 ]
机构
[1] Astra Hassle AB, Analyt Chem, S-43183 Molndal, Sweden
关键词
column liquid chromatography; glucoamylase; enantioselective retention and resolution; effect of operating parameters; ionic strength;
D O I
10.1007/BF02490654
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Several mobile phase parameters were investigated for controlling enantioselective retention and resolution on a chiral stationary phase made in-house. The chiral selector was the enzyme amyloglucosidase, which was immobilized onto a silica support via reductive amination. The influences of the mobile phase pH, concentration and type of uncharged organic modifier, ionic strength and column temperature on enantioselectivity were studied. The analysis time for resolving enantiomers could be adjusted with only a minor decrease in enantioselectivity by using a high ionic strength mobile phase buffer. This indicated a retention mechanism involving ion-exchange interactions. It was further confirmed by the decreasing enantioselectivity of amines when using a mobile phase pH below the isoelectric point of the native protein. Interesting effects were observed when the organic modifier concentration was increased and also when the column temperature was raised. Both retention and enantioselectivity increased with increasing concentration of 2-propanol in the mobile phase. Examples are given where both enantioselectivity and retention increased with increasing column temperature. Thermodynamic studies were performed to calculate the entropy and enthalpy constants. The results showed that, depending on mobile phase composition, the enantioselective retention may be caused by differences in entropy or enthalpy.
引用
收藏
页码:215 / 222
页数:8
相关论文
共 19 条
[1]   DIRECT LIQUID-CHROMATOGRAPHIC SEPARATION OF ENANTIOMERS ON IMMOBILIZED PROTEIN STATIONARY PHASES .3. OPTICAL RESOLUTION OF A SERIES OF N-AROYL D,L-AMINO ACIDS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ON BOVINE SERUM-ALBUMIN COVALENTLY BOUND TO SILICA [J].
ALLENMARK, S ;
BOMGREN, B ;
BOREN, H .
JOURNAL OF CHROMATOGRAPHY, 1983, 264 (01) :63-68
[2]  
ALLENMARK S, 1989, CHIRAL SEPARATIONS H, pCH11
[3]   MACROCYCLIC ANTIBIOTICS AS A NEW CLASS OF CHIRAL SELECTORS FOR LIQUID-CHROMATOGRAPHY [J].
ARMSTRONG, DW ;
TANG, YB ;
CHEN, SS ;
ZHOU, YW ;
BAGWILL, C ;
CHEN, JR .
ANALYTICAL CHEMISTRY, 1994, 66 (09) :1473-1484
[4]   RETENTION PROCESSES ON ALPHA-1-ACID GLYCOPROTEIN-BONDED STATIONARY PHASE [J].
ARVIDSSON, E ;
JANSSON, SO ;
SCHILL, G .
JOURNAL OF CHROMATOGRAPHY, 1992, 591 (1-2) :55-63
[5]  
DINGENEN J, 1994, PRACTICAL APPROACH C, P117
[6]   SEPARATION OF THE ENANTIOMERS OF BETA-RECEPTOR BLOCKING-AGENTS AND OTHER CATIONIC DRUGS USING A CHIRAL-AGP COLUMN - BINDING-PROPERTIES AND CHARACTERIZATION OF IMMOBILIZED ALPHA-1-ACID GLYCOPROTEIN [J].
ENQUIST, M ;
HERMANSSON, J .
JOURNAL OF CHROMATOGRAPHY, 1990, 519 (02) :285-298
[7]   SEPARATION OF ENANTIOMERS OF BENZODIAZEPINES ON THE CHIRAL-AGP COLUMN [J].
FITOS, I ;
VISY, J ;
SIMONYI, M ;
HERMANSSON, J .
JOURNAL OF CHROMATOGRAPHY A, 1995, 709 (02) :265-273
[8]   Thermodynamic study of an unusual chiral separation. Propranolol enantiomers on an immobilized cellulase [J].
Fornstedt, T ;
Sajonz, P ;
Guiochon, G .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1997, 119 (06) :1254-1264
[9]   OPTIMIZATION OF THE SEPARATION OF ENANTIOMERS OF BASIC DRUGS - RETENTION MECHANISMS AND DYNAMIC MODIFICATION OF THE CHIRAL BONDING PROPERTIES ON AN ALPHA(1)-ACID GLYCOPROTEIN COLUMN [J].
HERMANSSON, J ;
GRAHN, A .
JOURNAL OF CHROMATOGRAPHY A, 1995, 694 (01) :57-69
[10]   RELATIONSHIP BETWEEN ENANTIOSELECTIVITY AND SOLUTE STRUCTURE ON A CHIRAL ALPHA-1-ACID GLYCOPROTEIN COLUMN [J].
HERMANSSON, J ;
STROM, K ;
SANDBERG, R .
CHROMATOGRAPHIA, 1987, 24 :520-526