Explantation of INCOR Left Ventricular Assist Device After Myocardial Recovery

被引:12
作者
Komoda, Takeshi [1 ]
Komoda, Satsuki [1 ]
Dandel, Michael [1 ]
Weng, Yuguo [1 ]
Hetzer, Roland [1 ]
机构
[1] Deutsch Herzzentrum Berlin, Dept Cardiothorac & Vasc Surg, D-13353 Berlin, Germany
关键词
D O I
10.1111/j.1540-8191.2008.00665.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: Myocardial recovery in patients supported by the INCOR left ventricular assist device (LVAD; Berlin Heart GmbH, Berlin, Germany) is seen in 5% of patients on this device worldwide. We describe improved surgical techniques for INCOR LVAD explantation. Methods: The outcome of INCOR LVAD implantation at our center and the operative techniques of device explantation were studied. The patients weaned from the device were followed up. Results: Out of 121 patients supported by the device, five (4.1%) were weaned from the device, whereas 34 patients (28.1%) underwent heart transplantation. In explantation surgery, the inflow cannula was removed (one case) or remained in the left ventricle after occlusion with an inflow cannula plug, with transection of the inflow cannula at its curve (two cases) or without transection (two cases). When the inflow cannula was occluded without the support of cardiopulmonary bypass (two cases), operative time (180 min and 210 min) was shorter than that with other explantation procedures. After mean follow-up of 2.4 years (range two months-four years) after device explantation, all five patients are alive, have not required heart transplantation and are in New York Heart Association class I (one case) or class II (four cases). After weaning from the device, no cerebrovascular complication was observed in any of the five patients. Conclusions: There is a possibility of weaning after INCOR implantation and surgical techniques for the removal of the INCOR LVAD should be further developed. doi: 10.1111/j.1540-8191.2008.00665.x (J Card Surg 2008;23:642-647)
引用
收藏
页码:642 / 647
页数:6
相关论文
共 14 条
[1]  
*BERLINHEART AG, INCOR IMPL VENTR ASS
[2]   Left ventricular assist device and drug therapy for the reversal of heart failure [J].
Birks, Emma J. ;
Tansley, Patrick D. ;
Hardy, James ;
George, Robert S. ;
Bowles, Christopher T. ;
Burke, Margaret ;
Banner, Nicholas R. ;
Khaghani, Asghar ;
Yacoub, Magdi H. .
NEW ENGLAND JOURNAL OF MEDICINE, 2006, 355 (18) :1873-1884
[3]  
Borovetz H S, 1995, ASAIO J, V41, P42, DOI 10.1097/00002480-199501000-00008
[4]   Long-term results in patients with idiopathic dilated cardiomyopathy after weaning from left ventricular assist devices [J].
Dandel, M ;
Weng, YG ;
Siniawski, H ;
Potapov, E ;
Lehmkuhl, HB ;
Hetzer, R .
CIRCULATION, 2005, 112 (09) :I37-I45
[5]   Long-term follow-up of thoratec ventricular assist device bridge-to-recovery patients successfully removed from support after recovery of ventricular function [J].
Farrar, DJ ;
Holman, WR ;
McBride, LR ;
Kormos, RL ;
Icenogle, TB ;
Hendry, PJ ;
Moore, CH ;
Loisance, DY ;
El-Banayosy, A ;
Frazier, H .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2002, 21 (05) :516-521
[6]   Clinical experience with an implantable, intracardiac, continuous flow circulatory support device: Physiologic implications and their relationship to patient selection [J].
Frazier, OH ;
Myers, TJ ;
Westaby, S ;
Gregoric, ID .
ANNALS OF THORACIC SURGERY, 2004, 77 (01) :133-141
[7]   First experiences with a novel magnetically suspended axial flow left ventricular assist device [J].
Hetzer, R ;
Weng, YG ;
Potapov, EV ;
Pasic, M ;
Drews, T ;
Jurmann, M ;
Hennig, E ;
Müller, J .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2004, 25 (06) :964-970
[8]   Does continuous flow left ventricular assist device technology have a positive impact on outcome pretransplant and posttransplant? [J].
Klotz, Stefan ;
Stypmann, Joerg ;
Welp, Henryk ;
Schmid, Christof ;
Drees, Gabriele ;
Rukosujew, Andreas ;
Scheld, Hans H. .
ANNALS OF THORACIC SURGERY, 2006, 82 (05) :1774-1778
[9]   Low incidence of myocardial recovery after left ventricular assist device implantation in patients with chronic heart failure [J].
Mancini, DM ;
Beniaminovitz, A ;
Levin, H ;
Catanese, K ;
Flannery, M ;
DiTullio, M ;
Savin, S ;
Cordisco, ME ;
Rose, E ;
Oz, M .
CIRCULATION, 1998, 98 (22) :2383-2389
[10]   Clinical experience with the MicroMed DeBakey ventricular assist device [J].
Noon, GP ;
Morley, DL ;
Irwin, S ;
Abdelsayed, SV ;
Benkowski, RJ ;
Lynch, BE .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S133-S138