A magnetic bearing system for a continuous ventricular assist device

被引:13
作者
Baloh, MJ
Allaire, PE
Hilton, EF
Wei, NH
Baun, D
Flack, RD
Olson, DB
Bearnson, GB
Khanwilkar, PS
机构
[1] Univ Virginia, ROMAC, MANE, Dept Elect Engn, Charlottesville, VA 22903 USA
[2] Univ Virginia, Dept Mech Aerosp & Nucl Engn, Charlottesville, VA USA
[3] Univ Utah, Artificial Heart Lab, Salt Lake City, UT USA
[4] MedQuest Prod Inc, Salt Lake City, UT USA
关键词
D O I
10.1097/00002480-199909000-00016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This article describes a prototype continuous flow ventricular assist device (CFVAD3) supported in magnetic bearings. The VAD is a small centrifugal Four bladed pump, The pump's geometry is explained. The CFVAD3 is the first compact VAD completely supported in magnetic bearings. The magnetic bearings are composed of an inlet side actuator divided into eight pole sets, and an outlet side actuator, also divided into eight pole sets. The pump operating performance was tested and found to be within the design flow rate of up to 9 L/min, and head up to 170 mm Hg for human circulatory support. Magnetic bearing operation out of center positions under various operating orientations were measured and found to be <1/6 of the bearing clearance, well within specifications, The expected magnetic bearing power loss has been calculated at approximately 6.5 watts.
引用
收藏
页码:450 / 454
页数:5
相关论文
共 12 条
[1]   CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER [J].
AKAMATSU, T ;
NAKAZEKI, T ;
ITOH, H .
ARTIFICIAL ORGANS, 1992, 16 (03) :305-308
[2]  
ALLAIRE PE, 1996, TRIBIOL T, V39
[3]   Implantable centrifugal pump with hybrid magnetic bearings [J].
Bearnson, GB ;
Olsen, DB ;
Khanwilkar, PS ;
Long, JW ;
Sinnott, M ;
Kumar, A ;
Allaire, PE ;
Baloh, M ;
Decker, J .
ASAIO JOURNAL, 1998, 44 (05) :M733-M736
[4]   Development of a prototype magnetically suspended rotor ventricular assist device [J].
Bearnson, GB ;
Maslen, EH ;
Olsen, DB ;
Allaire, PE ;
Khanwilkar, PS ;
Long, JW ;
Kim, HC .
ASAIO JOURNAL, 1996, 42 (04) :275-281
[5]  
BELENGER J, 1983, ARTIF ORGANS, V1, P74
[6]  
Butler K C, 1992, ASAIO J, V38, pM296, DOI 10.1097/00002480-199207000-00041
[7]  
Hilton EF, 1997, ASAIO J, V43, pM598
[8]  
HOGNESS JR, 1991, ARTIFICIAL HEART PRO, P65
[9]   SYSTEM CONSIDERATIONS FAVORING ROTARY ARTIFICIAL HEARTS WITH BLOOD-IMMERSED BEARINGS [J].
JARVIK, RK .
ARTIFICIAL ORGANS, 1995, 19 (07) :565-570
[10]  
KEITH FJ, 1990, P 2 INT S MAGN BEAR