Implantable centrifugal pump with hybrid magnetic bearings

被引:19
作者
Bearnson, GB
Olsen, DB
Khanwilkar, PS
Long, JW
Sinnott, M
Kumar, A
Allaire, PE
Baloh, M
Decker, J
机构
[1] MedQuest Prod, Salt Lake City, UT 84103 USA
[2] Univ Utah, Artificial Heart Res Lab, Salt Lake City, UT USA
[3] Univ Virginia, Dept Mech Aerosp & Nucl Engn, Charlottesville, VA USA
关键词
D O I
10.1097/00002480-199809000-00088
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Test methods and results of in vitro assessment of a centrifugal pump with a magnetically suspended impeller are provided. In vitro blood tests have been completed with a resulting normalized milligram index of hemolysis (NmIH) of 12.4 +/- 4.1, indicating that hemolysis is not a problem. Hydraulic characterization of the system with water has shown that a nominal pumping condition of 6 L/min at 100 mmHg was met at 2,200 rpm. Maximum clinically usable cardiac output is predicted be 10 L/min. The magnetic bearing supported impeller did not contact the housing and was shown to be stable under a variety of pumping conditions. The driving motor efficiency is 75% at the nominal condition. Finally, a description of the clinical version of the pump under development is provided.
引用
收藏
页码:M733 / M736
页数:4
相关论文
共 15 条
[1]   Prototype continuous flow ventricular assist device supported on magnetic bearings [J].
Allaire, PE ;
Kim, HC ;
Maslen, EH ;
Olsen, DB ;
Bearnson, GB .
ARTIFICIAL ORGANS, 1996, 20 (06) :582-590
[2]   Development of a prototype magnetically suspended rotor ventricular assist device [J].
Bearnson, GB ;
Maslen, EH ;
Olsen, DB ;
Allaire, PE ;
Khanwilkar, PS ;
Long, JW ;
Kim, HC .
ASAIO JOURNAL, 1996, 42 (04) :275-281
[3]  
BLACKSHEAR PL, 1988, HDB BIOENGINEERING, P1
[4]  
DeBakey ME, 1997, ARTIF ORGANS, V21, P1149
[5]  
Hilton EF, 1997, ASAIO J, V43, pM598
[6]   Using hybrid magnetic bearings to completely suspend the impeller of a ventricular assist device [J].
Khanwilkar, P ;
Olsen, D ;
Bearnson, G ;
Allaire, P ;
Maslen, EA ;
Flack, R ;
Long, J .
ARTIFICIAL ORGANS, 1996, 20 (06) :597-604
[7]  
KUMAR BA, 1998, ASAIO J
[8]  
KUMAR BA, 1997, ASAIO J, V43, P2
[9]   Development of an implantable ventricular assist system [J].
Macris, MP ;
Parnis, SM ;
Frazier, OH ;
Fuqua, JM ;
Jarvik, RK .
ANNALS OF THORACIC SURGERY, 1997, 63 (02) :367-370
[10]  
MASON RG, 1982, HEMOSTASIS THROMBOSI, P944