摘要
Test methods and results of in vitro assessment of a centrifugal pump with a magnetically suspended impeller are provided. In vitro blood tests have been completed with a resulting normalized milligram index of hemolysis (NmIH) of 12.4 +/- 4.1, indicating that hemolysis is not a problem. Hydraulic characterization of the system with water has shown that a nominal pumping condition of 6 L/min at 100 mmHg was met at 2,200 rpm. Maximum clinically usable cardiac output is predicted be 10 L/min. The magnetic bearing supported impeller did not contact the housing and was shown to be stable under a variety of pumping conditions. The driving motor efficiency is 75% at the nominal condition. Finally, a description of the clinical version of the pump under development is provided.
引用
收藏
页码:M733 / M736
页数:4
相关论文
共 15 条
[11]  
Mizuno T., 1992, P 3 INT S MAGN BEAR
[12]  
Park CH, 1996, ARTIF ORGANS, V20, P128
[13]  
SALZMAN EW, 1987, HOMEOSTASIS THROMBOS
[14]   MINIMIZATION OF HEMOLYSIS IN CENTRIFUGAL BLOOD PUMPS - INFLUENCE OF DIFFERENT GEOMETRIES [J].
SCHIMA, H ;
MULLER, MR ;
PAPANTONIS, D ;
SCHLUSCHE, C ;
HUBER, L ;
SCHMIDT, C ;
TRUBEL, W ;
THOMA, H ;
LOSERT, U ;
WOLNER, E .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1993, 16 (07) :521-529
[15]   PLATELET SIZE DISTRIBUTION MEASUREMENTS AS INDICATORS OF SHEAR STRESS-INDUCED PLATELET-AGGREGATION [J].
SLACK, SM ;
JENNINGS, LK ;
TURITTO, VT .
ANNALS OF BIOMEDICAL ENGINEERING, 1994, 22 (06) :653-659