A computerised quality control testing system for B-mode ultrasound

被引:60
作者
Gibson, NM [1 ]
Dudley, NJ [1 ]
Griffith, K [1 ]
机构
[1] City Hosp Nottingham, Dept Med Phys, NHS Trust, Nottingham NG5 1PB, England
关键词
quality control; image processing; automated measurement; test object;
D O I
10.1016/S0301-5629(01)00479-3
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Current methods of ultrasound (US) imaging quality control involve an observer taking a series of measurements on test object images. This process is very labour-intensive, rapidly becoming more so as the complexity of US scanners increases. Also, many of the measurements are based on the subjective judgement of the operator and are, therefore, prone to an undesirable level of intraobserver and interobserver variation. We have developed a suite of programs to analyse captured US images to estimate a number of performance parameters, including resolution, high- and low-contrast penetration depths and high- and low-contrast sensitivity. In this system, images of the ATS Laboratories Multipurpose Phantom Model 539 are acquired on a PC with a video capture card, and analysed using the software developed. Automated tests have been developed for resolution, low- and high-contrast penetration and low- and high-contrast sensitivity. A preliminary validation of the system was carried out using images from two curvilinear probes and a linear-array probe operating at 5 MHz, 7.5 MHz and 10 MHz, respectively. The 95% confidence intervals ranged from +/- 17% to +/- 21% for resolution measurements and from 0% to +/- 1.5% for low-contrast penetration. The system provides an index of visibility for high- and low-contrast targets, as compared to the coarse visual assessment of a human observer, with 95% confidence intervals ranging from +/- 6% to +/- 13% for low-contrast sensitivity and from +/- 3% to +/- 25% for high-contrast sensitivity. Advantages of the system over the human observer include improved reproducibility and increased information regarding visibility of targets. A more detailed multicentre validation is currently being undertaken. (E-mail: ngibson@ncht.org.uk) (C) 2002 World Federation for Ultrasound in Medicine Biology.
引用
收藏
页码:1697 / 1711
页数:15
相关论文
共 12 条
[1]  
ALTMAN DG, 1991, PRACTICAL STAT MED R, P293
[2]  
[Anonymous], NOISE
[3]  
DUDLEY NJ, 2001, EUR J ULTRASOUND, V12, P223
[4]   ACCURATE ASSESSMENT OF PRECISION ERRORS - HOW TO MEASURE THE REPRODUCIBILITY OF BONE DENSITOMETRY TECHNIQUES [J].
GLUER, CC ;
BLAKE, G ;
LU, Y ;
BLUNT, BA ;
JERGAS, M ;
GENANT, HK .
OSTEOPOROSIS INTERNATIONAL, 1995, 5 (04) :262-270
[5]   Real-time B-mode ultrasound quality control test procedures - Report of AAPM ultrasound task group no. 1 [J].
Goodsitt, MM ;
Carson, PL ;
Witt, S ;
Hykes, DL ;
Kofler, JM .
MEDICAL PHYSICS, 1998, 25 (08) :1385-1406
[6]  
GRIFFITH K, 1995, BRIT J RADIOL, V68, P798
[7]  
KOFLER JM, 2000, J ULTRAS MED, V19, pS18
[8]  
LATHI BP, 1983, MODERN DIGITAL ANALO, P393
[9]  
LATHI BP, 1983, MODERN DIGITAL ANALO, P559
[10]  
PRICE R, 1995, 71 IPSM