Sorption of arsenic, bismuth, mercury, antimony, selenium and tin on dithiocarbamate loaded polyurethane foam as a preconcentration method for their determination in water samples by simultaneous inductively coupled plasma atomic emission spectrometry and electrothermal atomic absorption spectrometry

被引:126
作者
Arpadjan, S
Vuchkova, L
Kostadinova, E
机构
[1] University of Sofia, Faculty of Chemistry
关键词
column solid-phase extraction; trace element preconcentration water samples; inductively coupled plasma atomic emission spectrometry; electrothermal atomic absorption spectrometry;
D O I
10.1039/a606917g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Column solid-phase extraction using dithiocarbamate loaded polyurethane foam was applied to the preconcentration of trace amounts of As, Bi, Hg, Sb, Se and Sn from water samples prior to their measurement by simultaneous ICP-AES and ETAAS, The sorption recoveries of all the analytes were higher than 97% for 150 ml water sample solutions passed through the column with pH approximate to 4.5 and sodium chloride concentration up to 3%. For As, Bi, Hg, Sb and Se, quantitative solid-phase extraction can be achieved over a wide pH range, from 0.5 to pH 5. The combination of the proposed preconcentration method with subsequent simultaneous analyte determination in the methanol eluates by ICP-AES permits the detection of 3 mu g l(-1) As and Se, 8 mu g l(-1) Bi, 0.12 mu g l(-1) Hg, 2 mu g l(-1) Sb and 6 mu g l(-1) Sn in water samples. ETAAS measurement after dissolution of the sorbed analytes in isobutyl methyl ketone allows the detection of 0.06 mu g l(-1) As and Sb, 0.1 mu g l(-1) Bi and Sn, 0.08 mu g l(-1) Se and 0.3 mu g l(-1) Hg.
引用
收藏
页码:243 / 246
页数:4
相关论文
共 16 条
[1]   DETERMINATION OF TRACE-ELEMENTS IN ANALYTICAL-REAGENT GRADE SODIUM-SALTS BY ATOMIC-ABSORPTION SPECTROMETRY AND INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY AFTER PRECONCENTRATION BY COLUMN SOLID-PHASE EXTRACTION [J].
ALEXANDROVA, A ;
ARPADJAN, S .
ANALYST, 1993, 118 (10) :1309-1312
[2]   COLUMN SOLID-PHASE EXTRACTION AS PRECONCENTRATION METHOD FOR TRACE-ELEMENT DETERMINATION IN OXALIC-ACID BY ATOMIC-ABSORPTION SPECTROMETRY AND INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
ALEXANDROVA, A ;
ARPADJAN, S .
ANALYTICA CHIMICA ACTA, 1995, 307 (01) :71-77
[3]   TRENDS IN USING RESILIENT POLYURETHANE FOAMS AS SORBENTS IN ANALYTICAL-CHEMISTRY [J].
BRAUN, T .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1983, 314 (07) :652-656
[4]   CELLULOSE TETPA - A CHELATING COLLECTOR DESIGNED FOR MULTIELEMENT PRECONCENTRATION IN FLOW SYSTEMS [J].
BURBA, P ;
ROCHA, JC ;
SCHULTE, A .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 346 (04) :414-419
[5]   IMINODIACETIC ACID ETHYLCELLULOSE AS A CHELATING ION-EXCHANGER .1. DETERMINATION OF TRACE-METALS BY ATOMIC-ABSORPTION SPECTROMETRY AND COLLECTION OF URANIUM [J].
HORVATH, Z ;
LASZTITY, A ;
SZAKACS, O ;
BOZSAI, G .
ANALYTICA CHIMICA ACTA, 1985, 173 (JUL) :273-280
[7]   CHELATING POLYMERS AND RELATED SUPPORTS FOR SEPARATION AND PRECONCENTRATION OF TRACE-METALS [J].
KANTIPULY, C ;
KATRAGADDA, S ;
CHOW, A ;
GESSER, HD .
TALANTA, 1990, 37 (05) :491-517
[8]   EXTRACTION OF METALS FROM AQUEOUS-SOLUTION WITH POLYURETHANE FOAM [J].
MALONEY, MP ;
MOODY, GJ ;
THOMAS, JDR .
ANALYST, 1980, 105 (1256) :1087-1097
[9]  
PRAKASH N, 1989, MIKROCHIM ACTA, V3, P257
[10]   SEPARATION OF NOBLE-METAL IONS ON SILICA WITH CHEMICALLY BONDED LIGANDS [J].
SCHILLING, T ;
SCHRAMEL, P ;
MICHALKE, B ;
KNAPP, G .
MIKROCHIMICA ACTA, 1994, 116 (1-3) :83-90