Statistics of ultrasonic spectral parameters for prostate and liver examinations

被引:62
作者
Lizzi, FL
Feleppa, EJ
Astor, M
Kalisz, A
机构
[1] Biomedical Laboratories, Riverside Research Institute, New York
基金
美国国家卫生研究院;
关键词
Bandwidth - Medical imaging - Parameter estimation - Probability density function - Spectrum analysis - Statistics;
D O I
10.1109/58.655209
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A theoretical analysis was performed to describe statistical characteristics of calibrated spectral parameters used for ultrasonic tissue evaluation in the prostate and liver. The analysis assumes that radiofrequency (rf) echo signals exhibit Gaussian statistics. It derives the probability density function (pdf) of spectral parameters that are computed using sliding-window analysis techniques. The analysis relates the standard deviations of linear-regression spectral-parameter estimates to system and analysis parameters including bandwidth, center frequency, and the length of the sliding analysis window. The analysis also derives the pdf for mid-band fit parameter images. Theoretical results are found to agree well with clinical data from homogeneous segments in liver and prostate. The results offer a basis for evaluating spectral-estimator precision and for conducting future studies of lesion detectability based on spectral features.
引用
收藏
页码:935 / 942
页数:8
相关论文
共 24 条
[1]  
ABRAMOWITZ M, 1972, NBS APPLIED MATH SER, V55
[2]   Error bounds on ultrasonic scatterer size estimates [J].
Chaturvedi, P ;
Insana, MF .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1996, 100 (01) :392-399
[3]  
Davenport W. B., 1958, INTRO THEORY RANDOM, P158
[4]   DIAGNOSTIC SPECTRUM ANALYSIS IN OPHTHALMOLOGY - A PHYSICAL PERSPECTIVE [J].
FELEPPA, EJ ;
LIZZI, FL ;
COLEMAN, DJ ;
YAREMKO, MM .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1986, 12 (08) :623-631
[5]   Typing of prostate tissue by ultrasonic spectrum analysis [J].
Feleppa, EJ ;
Kalisz, A ;
SokilMelgar, JB ;
Lizzi, FL ;
Liu, T ;
Rosado, AL ;
Shao, MC ;
Fair, WR ;
Wang, Y ;
Cookson, MS ;
Reuter, VE ;
Heston, WDW .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (04) :609-619
[6]  
FELEPPA EJ, 1994, P IEEE ULTR S, P1483
[7]   Precision and accuracy of acoustospectrographic parameters [J].
Huisman, HJ ;
Thijssen, JM .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1996, 22 (07) :855-871
[8]   DESCRIBING SMALL-SCALE STRUCTURE IN RANDOM-MEDIA USING PULSE-ECHO ULTRASOUND [J].
INSANA, MF ;
WAGNER, RF ;
BROWN, DG ;
HALL, TJ .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 87 (01) :179-192
[9]  
KAPLAN D, 1994, P IEEE ULTR S, P961
[10]   VARIATION OF ACOUSTIC ATTENUATION COEFFICIENT SLOPE ESTIMATES FOR INVIVO LIVER [J].
KUC, R ;
TAYLOR, KJW .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1982, 8 (04) :403-412