Design and fabrication of a micromachined Coulter counter

被引:74
作者
Koch, M [1 ]
Evans, AGR [1 ]
Brunnschweiler, A [1 ]
机构
[1] Univ Southampton, Dept ECS, Southampton SO17 1BJ, Hants, England
关键词
D O I
10.1088/0960-1317/9/2/313
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper reports the design and fabrication of a micromachined Coulter counter. Calculations have been performed to estimate the behaviour of the counter for passing particles. A relative resistance change of 1.8% has been derived for a particle of 1.5 mu m radius flowing through a capillary of 5 mu m side length and an electrode spacing of 40 mu m. The fabrication technology has been based on similar micromachining steps to a micromixer, and allows further design modification to generate other microfluidic devices. Thus, integration of other microfluidic devices is possible with this technology. The fabrication relies on silicon trench etching and subsequent deposition of metal electrodes over the trench edges. Finally, a Pyrex wafer is anodically bonded on top of the silicon to seal the capillaries.
引用
收藏
页码:159 / 161
页数:3
相关论文
共 6 条
[1]  
Allen T., 1990, PARTICLE SIZE MEASUR, VFourth
[2]   MICROMACHINING OF CAPILLARY ELECTROPHORESIS INJECTORS AND SEPARATORS ON GLASS CHIPS AND EVALUATION OF FLOW AT CAPILLARY INTERSECTIONS [J].
FAN, ZH ;
HARRISON, DJ .
ANALYTICAL CHEMISTRY, 1994, 66 (01) :177-184
[3]   Two simple micromixers based on silicon [J].
Koch, M ;
Chatelain, D ;
Evans, AGR ;
Brunnschweiler, A .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1998, 8 (02) :123-126
[4]  
KOUTSOURIS D, 1988, BIORHEOLOGY, V25, P763
[5]  
LARSEN UD, 1997, P TRANSD 97 CHIC JUN, P1319, DOI DOI 10.1109/SENSOR.1997.635479
[6]   A SILICON MICROMACHINED DEVICE FOR USE IN BLOOD-CELL DEFORMABILITY STUDIES [J].
TRACEY, MC ;
GREENAWAY, RS ;
DAS, A ;
KAYE, PH ;
BARNES, AJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (08) :751-761