Measurement of the presampled two-dimensional modulation transfer function of digital imaging systems

被引:31
作者
Fetterly, KA [1 ]
Hangiandreou, NJ
Schueler, BA
Ritenour, ER
机构
[1] Mayo Clin, Dept Diagnost Radiol, Rochester, MN 55905 USA
[2] Univ Minnesota, Dept Radiol, Minneapolis, MN 55455 USA
关键词
two-dimensional modulation transfer function; computed radiography; digital radiography; detective quantum efficiency;
D O I
10.1118/1.1472498
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this work was to develop methods to measure the presampled two-dimensional modulation transfer function (2D MTF) of digital imaging systems. A custom x-ray "point source" phantom was created by machining 256 holes with diameter 0.107 mm through a 0.5-mm-thick copper plate. The phantom was imaged several times, resulting in many images of individual x-ray "spots." The center of each spot (with respect to the pixel matrix) was determined to subpixel accuracy by fitting each spot to a 2D Gaussian function. The subpixel spot center locations were used to create a 5X oversampled system point spread function (PSF), which characterizes the optical and electrical properties of the system and is independent of the pixel sampling of the original image. The modulus of the Fourier transform of the PSF was calculated. Next, the Fourier function was normalized to the zero frequency value. Finally, the Fourier transform function was divided by the first-order Bessel function that defined the frequency content of the holes. resulting in the presampled 2D MTF. The presampled 2D MTF of a 0.1 mm pixel pitch computed radiography system and 0.2 mm pixel pitch flat panel digital imaging system that utilized a cesium iodide scintillator was measured. Comparison of the axial components of the 2D MTF to one-dimensional MTF measurements acquired using an edge device method demonstrated that the two methods produced consistent results. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:913 / 921
页数:9
相关论文
共 23 条
[2]  
BATH M, 2001, P SPIE, V4320
[3]   Visual signal detection in structured backgrounds. III. Calculation of figures of merit for model observers in statistically nonstationary backgrounds [J].
Bochud, FO ;
Abbey, CK ;
Eckstein, MP .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2000, 17 (02) :193-205
[4]   Performance characteristics of a Kodak computed radiography system [J].
Bradford, CD ;
Peppler, WW ;
Dobbins, JT .
MEDICAL PHYSICS, 1999, 26 (01) :27-37
[5]   STATISTICALLY DEFINED BACKGROUND - PERFORMANCE OF A MODIFIED NONPREWHITENING OBSERVER MODEL [J].
BURGESS, AE .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1994, 11 (04) :1237-1242
[6]  
Cunningham I, 2000, HDB MED IMAGING, V1
[7]   A METHOD FOR MODULATION TRANSFER-FUNCTION DETERMINATION FROM EDGE PROFILES WITH CORRECTION FOR FINITE-ELEMENT DIFFERENTIATION [J].
CUNNINGHAM, IA ;
FENSTER, A .
MEDICAL PHYSICS, 1987, 14 (04) :533-537
[8]   SIGNAL AND NOISE IN MODULATION TRANSFER-FUNCTION DETERMINATIONS USING THE SLIT, WIRE, AND EDGE TECHNIQUES [J].
CUNNINGHAM, IA ;
REID, BK .
MEDICAL PHYSICS, 1992, 19 (04) :1037-1044
[9]   DQE(F) OF 4 GENERATIONS OF COMPUTED RADIOGRAPHY ACQUISITION DEVICES [J].
DOBBINS, JT ;
ERGUN, DL ;
RUTZ, L ;
HINSHAW, DA ;
BLUME, H ;
CLARK, DC .
MEDICAL PHYSICS, 1995, 22 (10) :1581-1593
[10]   Image quality evaluation of a desktop computed radiography system [J].
Fetterly, KA ;
Hangiandreou, NJ .
MEDICAL PHYSICS, 2000, 27 (12) :2669-2679