Image quality evaluation of a desktop computed radiography system
被引:28
作者:
Fetterly, KA
[1
]
Hangiandreou, NJ
[1
]
机构:
[1] Mayo Clin & Mayo Fdn, Rochester, MN 55905 USA
关键词:
computed radiography (CR);
modulation transfer function (MTF);
noise power spectrum (NPS);
detective quantum efficiency (DQE);
image quality;
D O I:
10.1118/1.1326449
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) of the Lumisys ACR-2000 desktop computed radiography (CR) reader were measured and compared to equivalent measurements acquired from a Fuji AC-3 CR system. The one-dimensional (1D) MTF was measured from an image of a sharp edge and the 1D NPS was derived from a 2D NPS measured from a uniform field exposure. The energy dependent ideal input signal to noise ratio of the incident x-ray beams was estimated using published x-ray spectra and attenuation coefficients. Measurements were acquired using Agfa, Fuji, and Kodak storage phosphor plates and it was concluded that use of the Fuji plates resulted in the highest system DQE for the ACR-2000. The DQE was measured using exposures of 0.10, 1.0, and 10.0 mR from 70 and 120 kVp x-ray beams filtered with aluminum. The DQE of the Lumisys ACR-2000 was lower than that of the Fuji AC-3. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)01112-3].