Experimental comparison of noise and resolution for 2k and 4k storage phosphor radiography systems

被引:118
作者
Flynn, MJ [1 ]
Samei, E
机构
[1] Henry Ford Hlth Syst, Dept Diagnost Radiol, Detroit, MI 48202 USA
[2] Univ Michigan, Dept Nucl Engn & Radiol Sci, Ann Arbor, MI 48109 USA
关键词
chest radiography; digital radiography; computed radiography (CR); storage phosphor radiography; image quality; noise power spectrum (NPS); modulation transfer function (MTF); detective quantum efficiency (DQE); resolution; noise;
D O I
10.1118/1.598656
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to compare the image quality for a digital storage phosphor system using 1760x2140 (2k) and 3520x4280 (4k) image arrays. Measurements were made on a chest radiography system (Fuji FCR-9501) with special provisions to be operated in both 2k (standard) and 4k (HQ) modes. Presampled modulation transfer functions (MTF) were measured using an edge method. Noise power spectra (NPS) were determined for different input exposures by two-dimensional Fourier analysis. These measures along with exposure measurements and an x-ray spectral model were used to determine the frequency-dependent detective quantum efficiency DQE (f) of the system for the 4k and the 2k modes. The magnitude of the NPS for the 4k mode was about 1/2 that of the 2k mode. A MTF value of 0.5 was found at 1.25 cycles/mm for the 4k system and 1.50 cycles/mm for the 2k system. The 4k images had an extended MTF of 0.1 at 4.5 cycles/mm in the plate-scan direction. Overall, the DQE (f) of the 4k mode was slightly better than that for the 2k mode by about 0.02 due primarily to its better noise characteristics. (C) 1999 American Association of Physicists in Medicine. [S0094-2405 (99)00408-3].
引用
收藏
页码:1612 / 1623
页数:12
相关论文
共 25 条
[1]  
Dainty JC., 1974, IMAGE SCI
[2]   EFFECTS OF UNDERSAMPLING ON THE PROPER INTERPRETATION OF MODULATION TRANSFER-FUNCTION, NOISE POWER SPECTRA, AND NOISE EQUIVALENT QUANTA OF DIGITAL IMAGING-SYSTEMS [J].
DOBBINS, JT .
MEDICAL PHYSICS, 1995, 22 (02) :171-181
[3]   DQE(F) OF 4 GENERATIONS OF COMPUTED RADIOGRAPHY ACQUISITION DEVICES [J].
DOBBINS, JT ;
ERGUN, DL ;
RUTZ, L ;
HINSHAW, DA ;
BLUME, H ;
CLARK, DC .
MEDICAL PHYSICS, 1995, 22 (10) :1581-1593
[4]  
DOBBINS JT, 1992, RADIOLOGY, V185, P159
[5]  
DOI K, 1986, ICRU, V41
[6]  
DOI K, 1982, MTFS WIENER SPECTRA, V1
[7]   A SIMPLE METHOD FOR DETERMINING THE MODULATION TRANSFER-FUNCTION IN DIGITAL RADIOGRAPHY [J].
FUJITA, H ;
TSAI, DY ;
ITOH, T ;
DOI, K ;
MORISHITA, J ;
UEDA, K ;
OHTSUKA, A .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (01) :34-39
[8]   BASIC IMAGING PROPERTIES OF A COMPUTED RADIOGRAPHIC SYSTEM WITH PHOTOSTIMULABLE PHOSPHORS [J].
FUJITA, H ;
UEDA, K ;
MORISHITA, J ;
FUJIKAWA, T ;
OHTSUKA, A ;
SAI, T .
MEDICAL PHYSICS, 1989, 16 (01) :52-59
[9]  
FUJITA H, 1989, SPIE, V1090, P263
[10]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .2. NOISE WIENER SPECTRUM [J].
GIGER, ML ;
DOI, K ;
METZ, CE .
MEDICAL PHYSICS, 1984, 11 (06) :797-805