Determination of residual carbon by inductively-coupled plasma optical emission spectrometry with axial and radial view configurations

被引:126
作者
Gouveia, ST
Silva, FV
Costa, LM
Nogueira, ARA
Nóbrega, JA
机构
[1] Univ Fed Sao Carlos, Dept Quim, BR-13560970 Sao Carlos, SP, Brazil
[2] Univ Fed Ceara, Fortaleza, Ceara, Brazil
[3] Embrapa Pecuaria Sudeste, Sao Carlos, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
residual carbon content; microwave-assisted digestion; inductively-coupled plasma optical emission spectrometry; radial and axial configurations;
D O I
10.1016/S0003-2670(01)01255-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work it was evaluated the performance of inductively-coupled plasma optical emission spectrometers (ICP-OESs) with axial and radial view configurations for residual carbon content (RCC) determination. The effects of carbon compound source (urea, L-cysteine, and glucose), sample medium, and internal standards on RCC determination were systematically evaluated. All measurements were carried out with two ICP spectrometers using the carbon atomic emission lines: 247.857 and 193.025 rim. The results obtained using axial and radial configurations showed that both the carbon source and the sample medium did not affect significantly the emission intensities. The sample medium only caused drastic influence when H2SO4 was employed probably due to transport interference that can be corrected employing Y as internal standard. The sensitivity attained using axial view ICP-OES was 20-fold better than that reached using radial view ICP-OES based on the slopes of the analytical curves at the most sensitive wavelength (193.025 mn). Using radial and axial ICP-OESs, high concentrations of Fe (> 100 mg l(-1)) interfered at 247.857 nm wavelength. An addition-recovery experiment was made by adding urea to an acid-digested sample and all recoveries were in the 100 +/- 5% range for axial and radial measurements. At this wavelength, R.S.D. <2.0% (n = 10) and detection limits of 33 and 34 mug ml(-1) C, were measured for ICP-OESs with radial and axial configurations, respectively. Biological samples were acid-digested using a closed-vessel microwave-assisted procedure and RCC was determined using both ICP-OES configurations. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:269 / 275
页数:7
相关论文
共 8 条
[1]   SIGNAL ENHANCEMENT OF ELEMENTS DUE TO THE PRESENCE OF CARBON-CONTAINING COMPOUNDS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
JAUNAULT, L ;
MAURAS, Y ;
MERMET, JM ;
DELAPORTE, T .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1497-1498
[2]   DETERMINATION OF ORGANIC-CARBON IN NATURAL-WATER WITH INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY AFTER EVAPORATION OF INORGANIC CARBON [J].
EMTERYD, O ;
ANDERSSON, B ;
WALLMARK, H .
MICROCHEMICAL JOURNAL, 1991, 43 (02) :87-93
[3]  
Kingston H.M. (Skip), 1997, MICROWAVE ENHANCED C
[4]  
Knapp G, 1998, ATOM SPECTROSC, V19, P220
[5]   DETERMINATION OF THE RESIDUAL CARBON CONTENT BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY AFTER DECOMPOSITION OF BIOLOGICAL SAMPLES [J].
KRUSHEVSKA, A ;
BARNES, RM ;
AMARASIRIWARADENA, CJ ;
FONER, H ;
MARTINES, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (06) :845-850
[6]   Combination of classical dry ashing with stripping voltammetry in trace element analysis of biological materials: Review of literature published after 1978 [J].
Mader, P ;
Szakova, J ;
Curdova, E .
TALANTA, 1996, 43 (04) :521-534
[7]  
Thomsen V, 2000, SPECTROSCOPY, V15, P33
[8]   RESIDUES FROM BIOLOGICAL-MATERIALS AFTER PRESSURE DECOMPOSITION WITH NITRIC-ACID .3. INFLUENCE OF REACTION-PRODUCTS ON INVERSE VOLTAMMETRIC ELEMENT DETERMINATION [J].
WURFELS, M ;
JACKWERTH, E ;
STOEPPLER, M .
ANALYTICA CHIMICA ACTA, 1989, 226 (01) :31-41